Nellcor™ Respiration Rate
ADD-24
Operator’s Manual Addendum
Unable to Calculate
Error
: The monitoring system is unable to calculate a respiration rate
value because the SpO
2
or pulse rate value is unavailable.
Resolution
:
•
Check for correct application and placement of sensor, as
described in the
Operator’s Manual
.
•
Check for other environmental and patient conditions that can
affect sensor performance, as described in the
Operator’s Manual
.
Outside Range
Status:
One of the following conditions has occurred:
•
The patient’s pulse rate is outside the range of 40 -170 beats per
minute.
•
The patient’s respiration rate is >0.5 of pulse rate.
Resolution
: Assess patient condition in conjunction with SpO
2
and
pulse rate readings from the monitoring system to determine appropri-
ate intervention. Use an alternate means of monitoring respiration rate.
RR Interference
Error
: The monitoring system is unable to calculate a respiration rate
value due to signal interference.
Resolution
:
•
Ensure that the patient is still and quiet.
•
Check for a documented history of atrial fibrillation or implanted
pacemaker.
•
Use an alternate means of monitoring respiration rate.
Check Patient
Status
: Respiration rate is not posted because a Low SpO
2
alarm con-
dition is present.
Resolution
: Take appropriate clinical actions in response to the Low
SpO
2
alarm.
Table 1-4.
Additional Alarm, Error, and Status Messages (Continued)
Message
Description