Troubleshooting
Operator’s Manual Addendum
ADD-23
1.7
Troubleshooting
Existing troubleshooting and error codes remain the same as those listed in the
Operator’s Manual
. Additional messages and errors related to the Respiration
Rate parameter follow.
Note:
The Respiration Rate parameter is approved for use in the range 4 to 40 breaths per
minute. When respiration rate falls below or exceeds alarm limits (user-configurable
within the approved range), a respiration rate alarm is indicated. If the monitoring
system detects a respiration rate below or above the 4 to 40 operating range, it
continues to report 4 or 40 respectively. These values may indicate that the parameter
has calculated a value outside the approved range.
Table 1-4.
Additional Alarm, Error, and Status Messages
Message
Description
Low Respiration Rate
Alarm
: The patient’s respiration rate is below the low alarm limit.
Resolution
: Assess patient condition in conjunction with SpO
2
and
pulse rate readings from the monitoring system to determine appropri-
ate intervention.
High Respiration Rate
Alarm
: The patient’s respiration rate is above the high alarm limit.
Resolution
: Assess patient condition in conjunction with SpO
2
and
pulse rate readings from the monitoring system to determine appropri-
ate intervention.
No RR Sensor Connected
Error
: A non-respiratory sensor is connected to the monitoring system.
Resolution
: Remove the sensor and replace it with a respiratory
sensor.
Calculating...
Status:
The monitoring system is attempting to calculate a respiration
rate value. This message appears for a brief time at the beginning of
each monitoring session.
Resolution
: No action required. The message disappears when calcu-
lation is complete. If a calculation cannot be made, the message
UNABLE TO CALCULATE appears.