Introduction
Operator’s Manual Addendum
ADD-5
patients with significantly irregular cardiac rhythms have not been
established. Use an alternate means of monitoring ventilatory status for
patients with significant cardiac dysrhythmia or an implanted pacemaker.
Caution:
Safety and effectiveness of Respiration Rate in pediatric and neonatal
patients have not been established.
Caution:
Safety and effectiveness of Respiration Rate in pregnant or lactating women
have not been established.
Caution:
Safety and effectiveness of Respiration Rate in patients on mechanical
ventilation have not been established.
Caution:
An SpO
2
alarm may be the first indication of hypoventilation.
Caution:
A low SpO
2
alarm limit of at least 90% is recommended when monitoring
patients on supplemental oxygen.
Caution:
Use of the Respiration Rate parameter does not change the need to set
threshold limits appropriate to the patient being monitored.
Caution:
Nellcor Respiration Rate performance has not been established during cases
of extreme motion (for example seizure) or poor perfusion (for example
shock). In cases of extreme motion or poor perfusion other forms of
respiration rate monitoring are recommended.