and ultraviolet lamp as our proven earlier design. The particulate filter is rated for 5 micron,
so will catch most particulates and many cell types.
The Covaris CH05 chiller and WCS 2.0
operate in combination to circulate the instrument’s
bath water through the WCS 2.0 ultraviolent (UV) sterilizer and particulate filter to ensure that
the water remains clean and free of algae growth for up to one month. The system is
designed to run continuously, maintaining water quality and relieving the need for daily water
changes in Covaris Adaptive Focused Acoustics (AFA™) systems.
Figure 1
– The Covaris Water Conditioning System
The WCS 2.0 incorporates an elapsed hours timer and maintenance due indicator for filter
and UV lamp replacement after one year of use. Filter and UV lamp replacement may be
performed by the user by following instructions included in this manual.
The system’s power is supplied by a small external converter that operates with any
worldwide AC power source (100 to 240 volts AC) and connects to a power jack on the rear
panel of the WCS 2.0.
Water Conditioning System block diagram