AFA ultraCUBE User Manual 

 

23 

 

Appendix B 

– User Guidelines and Troubleshooting 

 

 

Use only DI or distilled water in the acoustic bath. 

 

Daily water change is not required - the water is sterilized and filtered by the WCS 2.0. 

 

Check and top-off the water level prior to daily use, to make up for evaporation. 

 

Bleach the degas pump lines and bath monthly to remove any hidden algae colonies 

– even with 

the WCS 2.0 operating, some colonies may take hold in tubing or crevices. 

 

Energy Saver option:  When the ultrasonicator is not in use, place the CH05 in standby mode by 
pressing the “START/STOP” button (refer to CH05 manual for additional information).  In standby, 
the pump continues to operate circulating the bath water through the WCS 2.0, but the cooling 
circuitry is not energized. 

 

 

 

Problem 

Possible Cause 

Recommendation 

Water is dripping onto the 
laboratory bench 

Condensation from circulation of 
chilled water, leaks from a hose 
fitting 

If conditioning chilled water, 
change to room temperature 
water to test if dripping stops.  If 
not, open unit and visually check 
fittings 

No operation, no power 

No +12V supply to the WCS 2.0 

Verify that power pack is plugged 
into AC wall power and 
connected to the 12V jack on the 
rear panel of the WCS 2.0 

MAINTENANCE indicator 
is illuminated, and CHG 
lamp icon is blinking 

More than 9500 hours have 
elapsed since last filter and UV 
lamp replacement 

See Appendix A for filter and 
lamp replacement instructions 

MAINTENANCE indicator 
is illuminated, no CHG 
lamp icon 

UV lamp is not installed or is 
damaged. 

See Appendix A for lamp 
replacement instructions 
 

 
 
 
 

Appendix C 

– Bio-safety References 

 

 

Centers for Disease Control (CDC), Universal Precautions for Prevention of Transmission of HIV and Other 
Bloodborne Infections, published 1987, updated 1996 

 

Clinical and Laboratory Standards Institute (CLSI), GP17-A3 Clinical Laboratory Safety; Approved Guideline 
- Third Edition, published 2012, ISBN 1-56238-797-9 

 

Clinical and Laboratory Standards Institute (CLSI), M29-A4 Protection of Laboratory Workers from 
Occupationally Acquired Infections; Approved Guideline, Fourth Edition, published 2014, ISBN 1-56238-961-

 

Occupational Safety and Health Administration (OSHA), 29 CFR 1910.1030 Bloodborne Pathogens 

 

International Standards Organization (ISO) 15190:2003, Medical Laboratories 

– Requirements for Safety 

 

Содержание AFA ultraCUBE

Страница 1: ...traCUBE User Manual 1 AFA ultraCUBE Water Management System USER MANUAL A system developed for maintaining bath water quality in Covaris AFA instrumentation Product Name AFA ultraCUBE Product Number 5...

Страница 2: ...NIVERSAL PRECAUTIONS Universal Precautions should be followed on all specimen samples regardless of whether a sample is known to contain an infectious agent Laboratories handling specimen samples are...

Страница 3: ...D YOU HAVE ANY QUESTIONS CONCERNING THIS EQUIPMENT Warranty Services The purchased equipment is covered by a twelve 12 month warranty which includes all the service and support necessary so that the c...

Страница 4: ...AFA ultraCUBE User Manual 4 This page intentionally left blank...

Страница 5: ...ENTS 1 0 INTRODUCTION 6 2 0 ultraCUBE Installation and Setup 10 3 0 Indicators and Controls 12 4 0 System Operation 14 5 0 Specifications 15 6 0 Service 16 Appendix B User Guidelines and Troubleshooti...

Страница 6: ...enience if disregarded CAUTION Equipment damage may occur WARNING Personal injury may occur 1 2 Safety Symbols and Warnings The following symbols are employed in product labeling and in this manual Wh...

Страница 7: ...T operate the equipment and contact Covaris immediately 8 If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Do not remove...

Страница 8: ...cally switch between chilling heating mode and standby mode in which water continues to circulate through the WCS to maintain water quality without the energy consumption required for cooling System C...

Страница 9: ...continuously maintaining water quality and relieving the need for daily water changes in Covaris Adaptive Focused Acoustics AFA systems Figure 1 The Covaris Water Conditioning System The WCS 2 0 incor...

Страница 10: ...ermoCube manual on CD Setup Refer to Appendix A of this manual to install the quartz tube and UV lamp into the WCS 2 0 Place the WCS 2 0 and the CH05 on the bench next to the Ultrasonicator They may b...

Страница 11: ...AFA ultraCUBE User Manual 11...

Страница 12: ...or 60 Hz NOTE The WCS 2 0 unit is shipped with the UV chamber quartz tube and UV lamp not installed Install these items before powering the unit or using it in any way See Appendix A for UV lamp and...

Страница 13: ...ractical after the MAINTENANCE indicator illuminates The indicator is shut off by pressing an internal RESET button once replacement is performed Please see Appendix A for UV lamp and filter replaceme...

Страница 14: ...is is normal and the turbulence subsides once the WCS 2 0 fills with water The CH05 and WCS 2 0 system when operating will automatically circulate bath water filter it and retard the growth of algae o...

Страница 15: ...gh 33 cm 13 inches deep 28 cm 11 inches wide and 33 cm 13 inches high Weight Approximately 4 5 kilograms 10 lbs Approximately 12 7 kilograms 28 lbs Ambient Temp 19 C to 25 C 66 F to 77 C 0 C to 40 C 3...

Страница 16: ...ice covarisinc com CAUTION PREPARING THE WCS 2 0 FOR SHIPMENT PURGE THE WCS 2 0 OF WATER AND REMOVE THE UV LAMP QUARTZ TUBE Please purge the WCS 2 0 of standing water prior to return shipment particul...

Страница 17: ...2 0 500575 LE Series CH05 Hose Repair Kit 500576 LE Series WCS 2 0 Hose Repair Kit Quartz Tube Installation or Replacement The Ultraviolet UV chamber consists of the outer chamber in which a quartz tu...

Страница 18: ...and draw it out of the UV chamber by gently pulling upward There are two O rings which will likely come out with the tube If the quartz tube is being installed for the first time the two O rings will...

Страница 19: ...the water filter UV Lamp Enclosure Cap and Cable 5 Gently pull on the cable to withdraw the lamp socket from the enclosure If a lamp is already installed in the enclosure it will be drawn out with th...

Страница 20: ...ocessing of infectious or hazardous biological samples laboratory gloves should be worn and the used filter cartridge treated as biohazard waste in case any biological sample accidentally entered the...

Страница 21: ...ideways to release the fitting Repeat for the upper fitting The quick release fittings will seal and prevent water from draining from the cartridge or tubing 3 Pull on the used filter cartridge to rem...

Страница 22: ...ensure there are no water leaks 7 Once the system is verified to be operating correctly press and hold the RESET button located on the side of the electronics enclosure for about 5 seconds This action...

Страница 23: ...fittings No operation no power No 12V supply to the WCS 2 0 Verify that power pack is plugged into AC wall power and connected to the 12V jack on the rear panel of the WCS 2 0 MAINTENANCE indicator is...

Отзывы: