AFA ultraCUBE User Manual
23
Appendix B
– User Guidelines and Troubleshooting
Use only DI or distilled water in the acoustic bath.
Daily water change is not required - the water is sterilized and filtered by the WCS 2.0.
Check and top-off the water level prior to daily use, to make up for evaporation.
Bleach the degas pump lines and bath monthly to remove any hidden algae colonies
– even with
the WCS 2.0 operating, some colonies may take hold in tubing or crevices.
Energy Saver option: When the ultrasonicator is not in use, place the CH05 in standby mode by
pressing the “START/STOP” button (refer to CH05 manual for additional information). In standby,
the pump continues to operate circulating the bath water through the WCS 2.0, but the cooling
circuitry is not energized.
Problem
Possible Cause
Recommendation
Water is dripping onto the
laboratory bench
Condensation from circulation of
chilled water, leaks from a hose
fitting
If conditioning chilled water,
change to room temperature
water to test if dripping stops. If
not, open unit and visually check
fittings
No operation, no power
No +12V supply to the WCS 2.0
Verify that power pack is plugged
into AC wall power and
connected to the 12V jack on the
rear panel of the WCS 2.0
MAINTENANCE indicator
is illuminated, and CHG
lamp icon is blinking
More than 9500 hours have
elapsed since last filter and UV
lamp replacement
See Appendix A for filter and
lamp replacement instructions
MAINTENANCE indicator
is illuminated, no CHG
lamp icon
UV lamp is not installed or is
damaged.
See Appendix A for lamp
replacement instructions
Appendix C
– Bio-safety References
Centers for Disease Control (CDC), Universal Precautions for Prevention of Transmission of HIV and Other
Bloodborne Infections, published 1987, updated 1996
Clinical and Laboratory Standards Institute (CLSI), GP17-A3 Clinical Laboratory Safety; Approved Guideline
- Third Edition, published 2012, ISBN 1-56238-797-9
Clinical and Laboratory Standards Institute (CLSI), M29-A4 Protection of Laboratory Workers from
Occupationally Acquired Infections; Approved Guideline, Fourth Edition, published 2014, ISBN 1-56238-961-
0
Occupational Safety and Health Administration (OSHA), 29 CFR 1910.1030 Bloodborne Pathogens
International Standards Organization (ISO) 15190:2003, Medical Laboratories