background image

 

 
 
AFA ultraCUBE User Manual 

 

10 

 
 

 

2.0 

AFA ultraCUBE Installation and Setup 

 
Unpacking: 

 

Remove the CH05 from its box and remove the foam packaging. 

 

Remove the WCS 2.0 accessories and cardboard accessory tray from the WCS 2.0 
box. 

 

Pull the WCS 2.0 unit out of the box and remove side foam packaging. 

 
The WCS 2.0 box should contain the following items: 

 

This User’s Manual 

 

The WCS 2.0 unit 

 

12 volt “universal” input power supply module 

 

AC power line cord 

 

UV lamp (packaged separately in a cardboard tube to avoid shipping damage) 

 

Quartz tube for UV chamber (packaged separately to avoid shipping damage) 

 

One pair of laboratory gloves 

 

Water hoses to interconnect the CH05, WCS 2.0, and the Ultrasonicator 

 

 

The CH05 box should contain: 

 

The CH05 Chiller 

 

RS232 Interface cable for controlling the CH05 from the LE220-plus 

 

AC power line cord 

 

The ThermoCube document package, including the ThermoCube manual on CD. 

 

Setup: 

 

Refer to 

Appendix A

 of this manual to install the quartz tube and UV lamp into the 

WCS 2.0. 

 

 

Place the WCS 2.0 and the CH05 on the bench next to the Ultrasonicator.  They may 
be placed on either side, or on the floor to save bench space as long as the hoses 
reach from the front of the Ultrasonicator to the WCS 2.0. 

 

 

Connect the hoses between the WCS 2.0 and the CH05, using the two short hoses, 
as shown below. 

 

 

Connect the two longer hoses from the Ultrasonicator bath to the remaining ports on 
the WCS 2.0 

labeled “Water Tank”.  Either bath port can connect to either of the 

Water Tank WCS 2.0 ports.  Make certain that sufficient air flow is available for 
cooling the CH05.  

 
 

Note:

  Previous Covaris products used low pressure black neoprene hose.  Because the pump in 

the CH05 chiller can generate more pressure than the black hose can tolerate, the ultraCUBE 
comes with a white, high pressure hose.  Failure to use the correct hose can result in the hose 
failing and emptying the LE220-plus water bath.  See figure below to identify the correct hose 
type. 

Содержание AFA ultraCUBE

Страница 1: ...traCUBE User Manual 1 AFA ultraCUBE Water Management System USER MANUAL A system developed for maintaining bath water quality in Covaris AFA instrumentation Product Name AFA ultraCUBE Product Number 5...

Страница 2: ...NIVERSAL PRECAUTIONS Universal Precautions should be followed on all specimen samples regardless of whether a sample is known to contain an infectious agent Laboratories handling specimen samples are...

Страница 3: ...D YOU HAVE ANY QUESTIONS CONCERNING THIS EQUIPMENT Warranty Services The purchased equipment is covered by a twelve 12 month warranty which includes all the service and support necessary so that the c...

Страница 4: ...AFA ultraCUBE User Manual 4 This page intentionally left blank...

Страница 5: ...ENTS 1 0 INTRODUCTION 6 2 0 ultraCUBE Installation and Setup 10 3 0 Indicators and Controls 12 4 0 System Operation 14 5 0 Specifications 15 6 0 Service 16 Appendix B User Guidelines and Troubleshooti...

Страница 6: ...enience if disregarded CAUTION Equipment damage may occur WARNING Personal injury may occur 1 2 Safety Symbols and Warnings The following symbols are employed in product labeling and in this manual Wh...

Страница 7: ...T operate the equipment and contact Covaris immediately 8 If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Do not remove...

Страница 8: ...cally switch between chilling heating mode and standby mode in which water continues to circulate through the WCS to maintain water quality without the energy consumption required for cooling System C...

Страница 9: ...continuously maintaining water quality and relieving the need for daily water changes in Covaris Adaptive Focused Acoustics AFA systems Figure 1 The Covaris Water Conditioning System The WCS 2 0 incor...

Страница 10: ...ermoCube manual on CD Setup Refer to Appendix A of this manual to install the quartz tube and UV lamp into the WCS 2 0 Place the WCS 2 0 and the CH05 on the bench next to the Ultrasonicator They may b...

Страница 11: ...AFA ultraCUBE User Manual 11...

Страница 12: ...or 60 Hz NOTE The WCS 2 0 unit is shipped with the UV chamber quartz tube and UV lamp not installed Install these items before powering the unit or using it in any way See Appendix A for UV lamp and...

Страница 13: ...ractical after the MAINTENANCE indicator illuminates The indicator is shut off by pressing an internal RESET button once replacement is performed Please see Appendix A for UV lamp and filter replaceme...

Страница 14: ...is is normal and the turbulence subsides once the WCS 2 0 fills with water The CH05 and WCS 2 0 system when operating will automatically circulate bath water filter it and retard the growth of algae o...

Страница 15: ...gh 33 cm 13 inches deep 28 cm 11 inches wide and 33 cm 13 inches high Weight Approximately 4 5 kilograms 10 lbs Approximately 12 7 kilograms 28 lbs Ambient Temp 19 C to 25 C 66 F to 77 C 0 C to 40 C 3...

Страница 16: ...ice covarisinc com CAUTION PREPARING THE WCS 2 0 FOR SHIPMENT PURGE THE WCS 2 0 OF WATER AND REMOVE THE UV LAMP QUARTZ TUBE Please purge the WCS 2 0 of standing water prior to return shipment particul...

Страница 17: ...2 0 500575 LE Series CH05 Hose Repair Kit 500576 LE Series WCS 2 0 Hose Repair Kit Quartz Tube Installation or Replacement The Ultraviolet UV chamber consists of the outer chamber in which a quartz tu...

Страница 18: ...and draw it out of the UV chamber by gently pulling upward There are two O rings which will likely come out with the tube If the quartz tube is being installed for the first time the two O rings will...

Страница 19: ...the water filter UV Lamp Enclosure Cap and Cable 5 Gently pull on the cable to withdraw the lamp socket from the enclosure If a lamp is already installed in the enclosure it will be drawn out with th...

Страница 20: ...ocessing of infectious or hazardous biological samples laboratory gloves should be worn and the used filter cartridge treated as biohazard waste in case any biological sample accidentally entered the...

Страница 21: ...ideways to release the fitting Repeat for the upper fitting The quick release fittings will seal and prevent water from draining from the cartridge or tubing 3 Pull on the used filter cartridge to rem...

Страница 22: ...ensure there are no water leaks 7 Once the system is verified to be operating correctly press and hold the RESET button located on the side of the electronics enclosure for about 5 seconds This action...

Страница 23: ...fittings No operation no power No 12V supply to the WCS 2 0 Verify that power pack is plugged into AC wall power and connected to the 12V jack on the rear panel of the WCS 2 0 MAINTENANCE indicator is...