background image

 

 
 
AFA ultraCUBE User Manual 

 

18 

 
 

3.  Carefully remove the outside cover of the WCS 2.0 by pulling it up and off the WCS 

2.0 chassis. 

4.  If present, remove the UV lamp.  See instructions below. 
5.  Remove the white cap at the top of the UV chamber by turning it counter-clockwise. 

 

 

 

6.  If a quartz tube is present, grasp the top of the tube and draw it out of the UV 

chamber by gently pulling upward. There are two O-rings which will likely come out 
with the tube.  If the quartz tube is being installed for the first time, the two O-rings 
will be just below the white cap.  Reach one finger into the chamber and lift out the 
two O-rings. 

7.  With the two O-rings positioned about a half-inch from the top of the quartz tube, 

slide the tube into the UV chamber. 

        

 

 
8.  Thread the white cap onto the chamber by turning clockwise until snug.  Hand tighten 

only. 

9.  Reinstall UV lamp.  

 

 

 

 

 

 

 

Содержание AFA ultraCUBE

Страница 1: ...traCUBE User Manual 1 AFA ultraCUBE Water Management System USER MANUAL A system developed for maintaining bath water quality in Covaris AFA instrumentation Product Name AFA ultraCUBE Product Number 5...

Страница 2: ...NIVERSAL PRECAUTIONS Universal Precautions should be followed on all specimen samples regardless of whether a sample is known to contain an infectious agent Laboratories handling specimen samples are...

Страница 3: ...D YOU HAVE ANY QUESTIONS CONCERNING THIS EQUIPMENT Warranty Services The purchased equipment is covered by a twelve 12 month warranty which includes all the service and support necessary so that the c...

Страница 4: ...AFA ultraCUBE User Manual 4 This page intentionally left blank...

Страница 5: ...ENTS 1 0 INTRODUCTION 6 2 0 ultraCUBE Installation and Setup 10 3 0 Indicators and Controls 12 4 0 System Operation 14 5 0 Specifications 15 6 0 Service 16 Appendix B User Guidelines and Troubleshooti...

Страница 6: ...enience if disregarded CAUTION Equipment damage may occur WARNING Personal injury may occur 1 2 Safety Symbols and Warnings The following symbols are employed in product labeling and in this manual Wh...

Страница 7: ...T operate the equipment and contact Covaris immediately 8 If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Do not remove...

Страница 8: ...cally switch between chilling heating mode and standby mode in which water continues to circulate through the WCS to maintain water quality without the energy consumption required for cooling System C...

Страница 9: ...continuously maintaining water quality and relieving the need for daily water changes in Covaris Adaptive Focused Acoustics AFA systems Figure 1 The Covaris Water Conditioning System The WCS 2 0 incor...

Страница 10: ...ermoCube manual on CD Setup Refer to Appendix A of this manual to install the quartz tube and UV lamp into the WCS 2 0 Place the WCS 2 0 and the CH05 on the bench next to the Ultrasonicator They may b...

Страница 11: ...AFA ultraCUBE User Manual 11...

Страница 12: ...or 60 Hz NOTE The WCS 2 0 unit is shipped with the UV chamber quartz tube and UV lamp not installed Install these items before powering the unit or using it in any way See Appendix A for UV lamp and...

Страница 13: ...ractical after the MAINTENANCE indicator illuminates The indicator is shut off by pressing an internal RESET button once replacement is performed Please see Appendix A for UV lamp and filter replaceme...

Страница 14: ...is is normal and the turbulence subsides once the WCS 2 0 fills with water The CH05 and WCS 2 0 system when operating will automatically circulate bath water filter it and retard the growth of algae o...

Страница 15: ...gh 33 cm 13 inches deep 28 cm 11 inches wide and 33 cm 13 inches high Weight Approximately 4 5 kilograms 10 lbs Approximately 12 7 kilograms 28 lbs Ambient Temp 19 C to 25 C 66 F to 77 C 0 C to 40 C 3...

Страница 16: ...ice covarisinc com CAUTION PREPARING THE WCS 2 0 FOR SHIPMENT PURGE THE WCS 2 0 OF WATER AND REMOVE THE UV LAMP QUARTZ TUBE Please purge the WCS 2 0 of standing water prior to return shipment particul...

Страница 17: ...2 0 500575 LE Series CH05 Hose Repair Kit 500576 LE Series WCS 2 0 Hose Repair Kit Quartz Tube Installation or Replacement The Ultraviolet UV chamber consists of the outer chamber in which a quartz tu...

Страница 18: ...and draw it out of the UV chamber by gently pulling upward There are two O rings which will likely come out with the tube If the quartz tube is being installed for the first time the two O rings will...

Страница 19: ...the water filter UV Lamp Enclosure Cap and Cable 5 Gently pull on the cable to withdraw the lamp socket from the enclosure If a lamp is already installed in the enclosure it will be drawn out with th...

Страница 20: ...ocessing of infectious or hazardous biological samples laboratory gloves should be worn and the used filter cartridge treated as biohazard waste in case any biological sample accidentally entered the...

Страница 21: ...ideways to release the fitting Repeat for the upper fitting The quick release fittings will seal and prevent water from draining from the cartridge or tubing 3 Pull on the used filter cartridge to rem...

Страница 22: ...ensure there are no water leaks 7 Once the system is verified to be operating correctly press and hold the RESET button located on the side of the electronics enclosure for about 5 seconds This action...

Страница 23: ...fittings No operation no power No 12V supply to the WCS 2 0 Verify that power pack is plugged into AC wall power and connected to the 12V jack on the rear panel of the WCS 2 0 MAINTENANCE indicator is...

Отзывы: