- 17 - 

  

Consumable Parts, Replacement Parts, and Options 

 

 

Part Name 

Model 

Part Code 

Remarks 

Filter Element (10pcs) 

FE-2 

59160022 

 

Drain Filter 

DF-114 

59070018 

 

Inspection Gas 

EG-10D 

59150300 

 

Attachment 

AT-2B 

20524489 

For tubing and coupling 

Instruction Manual 

 

 

 

 

 

 

 

- 16 - 

 

6. Maintenance Check 

 

 

The Gas Leak Detector is precision measuring equipment. Maintenance in accordance 

with the following items is imperative for maintaining performance of the Gas Leak 

Detector and ensuring safety.   

In case of dropping the Gas Leak Detector or being wet by using under any conditions 

other than the normal environment (refer to 

Cautions 

in

 1. Introduction

 and 

Range

 in

 9. 

Specifications

) perform a daily inspection or request a periodic inspection (will be 

charged). 

 

 

 

  

Daily Check 

 

Items to Check 

Description 

Alarm Function 

Attach the Drain Filter and turn ON the power. After the warming up 

operation, set the gas sensitivity level to “3” and confirm that the 

alarm lamp (red) flashes and the alarm buzzer will sound by 

suctioning inspection gas for half a second. 

If the alarm sound indicator is displayed on the LCD and the alarm 

lamp (red) will not flash or the alarm buzzer will not sound, contact 

your New Cosmos distributor. 

Attachment 

Check the Attachment for any wear or damage which may affect 

operations.

 

If any, replace with a new Attachment. 

Filter Element 

When the Drain Filter is dirty or wet, or the water or liquid is in the 

Drain Filter, clean out the Drain Filter and replace the Filter Element.

 

(Refer to 

Filter Element Replacement

 on page 14.) 

Battery Charge 

If the remaining battery charge is low, replace the batteries.

 

(Refer 

to

 Battery Replacement

 on page 15.) 

Airtightness 

Turn ON the power and confirm that "E-P" (refer to 

Pump Error

 on 

page 13) is displayed on the LCD by blocking the inlet of the 

Attachment.If an error occurs, the airtightness is normal. Press the 

restart pump button. 

If an error does not occur, the Drain Filter is damaged or the airtight 

of diaphragm is damaged. Replace the parts or request repairs. 

 

 

 

  

Periodic Inspection 

 

To maintain the accuracy of the Gas Leak Detector, request a periodic inspection at your 

New Cosmos distributor or directly from New Cosmos at least once a year. 

Содержание XP-703D III

Страница 1: ...tector This instruction manual explains the use of three models listed on the left Keep this Instruction Manual available for quick reference when needed Read this Instruction Manual and thoroughly un...

Страница 2: ...ilent Mode No Operation Sound 11 Deactivating the Alarm Buzzer 12 Lighting the Backlight 12 4 Error Displays 13 5 Replacement of Consumable Parts 14 Filter Element Replacement 14 Battery Replacement 1...

Страница 3: ...in preparing the Gas Leak Detector However if any items are missing or damaged please contact your New Cosmos distributor Replacements for any missing or damaged items will be sent Name Quantity Gas...

Страница 4: ...stand thoroughly this Instruction Manual regardless of your experience with using Gas Leak Detectors Do not use this Gas Leak Detector for improper purpose Do not use this Gas Leak Detector in undocum...

Страница 5: ...the batteries are leaking when the remaining battery charge is low Do not disassemble alter or change the structure or electric circuits of the Gas Leak Detector It may affect its performance Do not l...

Страница 6: ...4 2 Part Names and Functions 1 2 10 6 8 4 3 9 7 5 11...

Страница 7: ...s light if the concentration is very high These lamps also flash when an error occurs in the pump or sensor 8 Buzzer hole The buzzer sounds 9 LCD Displays the gas sensitivity and messages a The remain...

Страница 8: ...6 No Name Functions 1 Side Zipper Open the zipper to replace the batteries 2 Back Pocket The Operating Guide is stored here 3 Inside Pocket A Replacement Filter Element is stored here Soft case 1 3 2...

Страница 9: ...d close the zippers both sides and Velcro tape The Gas Leak Detector is in the soft case from the shipment 2 Insert the Attachment to the Drain Filter The attachment is inserted to the drain filter fr...

Страница 10: ...p is completed the highest sensitivity level 4 lamp will flash The maximum warm up time is three minutes Always perform the daily check before use Refer to page 16 WARNING STEPS 3 Detection Turn OFF p...

Страница 11: ...as sensitivity lamp When the gas sensitivity level is decreased beyond 1 the level will return to level 4 Increasing the Gas Sensitivity Level Each time pressing the POWER SENSITIVITY button twice dou...

Страница 12: ...p sound for 3 seconds and then it will return to display the gas sensitivity level If Zero Adjustment is not able to be performed the buzzer sound will beep and it remains to display the gas sensitivi...

Страница 13: ...ncentration gas of any in the following table is detected leave the Gas Leak Detector in clean air for a certain period of time before turning OFF the power Turning OFF the power immediately may affec...

Страница 14: ...d ON OFF The alarm sound indicator on the LCD will light buzzer ON or go out buzzer OFF Lighting the Backlight The LCD backlight lights by pressing the ZERO RESTART PUMP BACKLIGHT button It also light...

Страница 15: ...g ON the power several times there may be a problem with the sensor Request repairs Pump Error Error lamps flash Water or liquid may have been sampled or the inlet may be blocked Remove the water or l...

Страница 16: ...function properly and will not be able to detect correctly Attach the filter case securely If the filter case is loose water or other foreign matter may leak into the Gas Leak Detector If water or li...

Страница 17: ...ipper 2 Open the battery lid of the back of the Gas Leak Detector and install two new batteries size AA alkaline as shown on the bottom of the compartment 3 Close the battery lid firmly 4 Zip the righ...

Страница 18: ...flash or the alarm buzzer will not sound contact your New Cosmos distributor Attachment Check the Attachment for any wear or damage which may affect operations If any replace with a new Attachment Fil...

Страница 19: ...lacement Parts and Options Part Name Model Part Code Remarks Filter Element 10pcs FE 2 59160022 Drain Filter DF 114 59070018 Inspection Gas EG 10D 59150300 Attachment AT 2B 20524489 For tubing and cou...

Страница 20: ...ll not turn ON by pressing the POWER SENSITIVITY button The battery polarity is reversed Insert the batteries correctly Battery Replacement on page 15 Battery exhaustion Replace the batteries Displayi...

Страница 21: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation...

Страница 22: ...ressure Range Atmospheric pressure 800 to 1100hPa Operating Temperature and Humidity Range 0 to 40 degrees C less than 85 RH with no condensation Dimensions Approx 38 W x 130 H x 32 D mm excluding pro...

Страница 23: ...of a metal oxide semiconductor that is sintered in a ball on a platinum wire coil that functions as both a heater and electrode It detects the change of electrical conductivity occurred by oxidation r...

Страница 24: ...naka Yodogawa ku Osaka 532 0036 Japan Tokyo Branch New Cosmos Electric Co Ltd 2 6 2 Hamamatsu cho Minato ku Tokyo 105 0013 Japan Phone 81 3 5403 2715 Fax 81 3 5403 2710 Email e info new cosmos co jp U...

Отзывы: