INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
14.1.2.1 Rinsing of Humidification System Procedure:
1.
Turn off the gas flow.
2. Empty the gas washing bottles.
3. Pour approximately 40 ml 70% alcohol into each gas washing bottle.
4. Set the gas washing bottles back into the gassing station.
5. Remove the gas dispersion nozzles using a small piece of tape as shown.
6.
Turn on the gas flow to approximately 30 l/h and leave the flow running for one hour.
7. Take some tissue paper and clean the gassing hole within the table plate with 70% alcohol while
the gas is flowing.
8.
After an hour, turn the gas flow off and empty the remains of the 70% alcohol left in the bottles.
9. Rinse the bottles with distilled water to ensure complete removal of the alcohol from the
humidification system.
10. Fill the gas washing bottles to 2/3 with distilled water.
11.
Re-establish gas flow to the system at 20-25 l/h and leave it running for 20 minutes.
12.
Turn OFF the gas flow.
13.
Refitting the gas dispersion nozzles to the table plate complete the cleaning procedure.