67
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
13 TROUBLESHOOTING
Heating System
SYMPTOM
CAUSE
ACTION
No heating, display is OFF
The heating switch is OFF
Press the heating switch.
No heating
The alarm is ON
The temperature is more than
0.5 °C off the set temperature.
No heating.
Set-point is too low
Raise the Set-point
No heat on one side, display
flashing off
Heat area is OFF
Turn on the heat area by
pressing HEAT and the DOWN
key at the same time.
Laminar Flow
SYMPTOM
CAUSE
ACTION
Alarm when the fan turns on
HEPA filter needs to be checked
Contact your Service
Representative for details.
Humidification System
SYMPTOM
CAUSE
ACTION
No Gas Flow
Gas washing Bottle is empty
Re-fill the Gas Washing Bottle.
Check and tighten gas
connections on the back of the
Workstation.
Loose tube connection
Check and tighten the bottle
connections.
Blocked gas dispersion nozzle
Remove and clean the
dispersion nozzles.
Nothing on Display
Loss of power
Check power cord.
Check fuses.