58
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
When all the average values are calculated, they should be compared with what is being shown by the
system. This is done by entering the menu and checking the values. Adjust only if the calculated values
are different than the ones displayed by the system for each zone.
NOTE:
Refer to sub-section 6.11.2 through 6.11.7 for more information regarding the calibration
functions for the tabletop.
CAUTION
Extreme care must be taken here, as any wrong changes or adjustments will
have a direct affect on the surface temperature of the tabletop.
9.2 Calibrating of Heated Glass
The heated glass is calibrated in the same way as the tabletop; except there is only one measurement
location taken in the middle of the glass.
NOTE:
Refer to sub-sections 6.11.9 and 6.11.10 for more information regarding the calibration functions
for the heated glass.