Follow these steps for a detailed instruction of how to validate/calibrate the temperature of the tabletop.
1. Switch on the workstation (
) and the fan (
). Then wait for at least 40 minutes
for the temperature to stabilize at set -point (e.g. 37.0 °C).
2. Place the dish containing oil-based liquid
and temperature sensor on the tabletop.
3. Wait for the temperature to stabilize. When
temperature is stabilized, write down the
value and move on to the next spot within
the same zone.
4. When the temperature of both spots has been measured, the average value must
be calculated using the following formula, where “x” is the number of the zone:
Temp
_
avr
tn-x
=
Temp
1
tn-x
+
Temp
2
tn-x
2
5. If there are more zones, repeat steps 2 through 4 for each zone.