56
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
9.1 Calibrating Temperature on the Tabletop
To locate the zones on the workstation, see Section 7: Different Heated Surfaces. In this section, the
temperature validation of the tabletop is explained in more detail. For each zone, two measurements
are taken on either side of the center of the zone. An example of the correct measurement location can
be seen in Figure 9-1, this is for a single T845-2 Right Side.
Figure 9-1: Single T845-2 Right Side with measurement spots.
NOTE:
Always measure the temperature on the whole tabletop before any adjusting is conducted.
Refusing to do so will cause heat crossover to other zones and make it very difficult to adjust
the temperature.
Содержание K-Systems L100 IVF Series
Страница 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Страница 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Страница 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Страница 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Страница 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Страница 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...