![Cooper Surgical LEEP PRECISION LP-30-120 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/cooper-surgical/leep-precision-lp-30-120/leep-precision-lp-30-120_operating-manual_2664728006.webp)
4
DISINFECTION
7 . Use a 70% IPA wipe, or a low-lint wipe, saturated with 70% IPA, to
thoroughly wipe the device. Pay particular attention to difficult to reach
areas such as device seams and area between buttons .
8 . With a fresh wipe saturated with 70% IPA, thoroughly wipe the device .
Allow the wiped and wetted device to stand for a minimum of the
disinfectant product manufacture’s labeled contact time . Re-wet the
wipes as necessary to ensure that all surfaces remain wet for the
entire time .
9 . Allow the device to air dry . Once dry, visually inspect the device for
any residual soil . Repeat cleaning/disinfection instructions if visible
soil is present .
10 . Place the disinfected device in an appropriate dry storage area .
Specifications
Use
Environmental Temperature:
b10 °C and +40 °C
Relative Humidity:
between 10 % and 90 %
Air Pressure:
between 700 hPa and 1060 hPa
Shipping and Storage
Environmental Temperature:
b10 °C and +40 °C
Relative Humidity:
between 10 % and 90 %
Air Pressure:
between 700 hPa and 1060 hPa
Содержание LEEP PRECISION LP-30-120
Страница 1: ...LEEP PRECISIONTM Smoke Evacuator 120 VAC Operating Manual REF LP 30 120 ...
Страница 15: ...13 NOTES ...