For procedures requiring open area plume removal (i.e., external lesions)
1 . Assemble the sterile Disposable Evacuation Tubing (Catalog number
6084) directly into the top of the Pre-Filter . Position the opposite end
over the site to be treated .
2 . Plug the LEEP PRECISION Smoke Evacuator into an appropriate wall
outlet (see Specifications) .
3 . Push the POWER switch to the ON position to start the unit .
4 . At the completion of each procedure, activate the system to ensure
safe particle containment . Using gloves and a mask, remove the
Pre-Filter, the Reducer, and the used section of Suction Tubing
and discard into an infectious waste receptacle (see Cautions) . The
CooperSurgical LEEP PRECISION Smoke Evacuator should be stored
with a new Pre-Filter and Reducer in place on the ULPA filter .
NOTE:
The expected life of the ULPA filter is 3–6 months depending on
usage . It must be discarded into an infectious waste receptacle if plume
odor is detected or the suction is diminished .
Maintenance
No maintenance is required during normal operation other than making
sure that ample space is maintained around the unit to allow for free air
flow and adequate cooling .
Cleaning and Disinfection
After each patient use, follow the combined cleaning/disinfection
instructions provided below for the LEEP PRECISION Smoke Evacuator .
CLEANING
1 . Prepare an enzymatic, neutral pH cleaner solution according to the
Manufacturer’s instructions .
2 . Immerse a clean, lint-free wipe into cleaning solution and thoroughly
wring . Thoroughly wipe the front panel surfaces in a circular motion .
Discard the wipe .
3 . Using a soft, nylon brush, scrub difficult-to-access areas such as
crevices or textured surfaces .
4 . Immerse a fresh, clean, lint-free wipe under warm flowing, utility water
and thoroughly wring excess water . Thoroughly wipe the front panel
surface and crevices for at least 30 seconds . Discard the wipe .
5 . Rinse a fresh, clean, lint-free wipe under warm flowing, utility water and
thoroughly wring excess water . Thoroughly wipe the front panel surface
and crevices for at least 30 seconds . Discard the wipe .
6 . Inspect the device for visible soil or debris . If visible soil remains,
repeat cleaning steps 2-5 until device is visually clean .