Device Description /Intended Use

The CooperSurgical LEEP PRECISION

 Smoke Evacuator is  

a stand alone, three-stage air filtration device that is used to capture 

airborne particulate from smoke plumes produced during electrosurgical 

procedures .  

Cautions

 

  

U .S . Federal law restricts this device to sale by or on the order  

of a physician .

 

   This device produces a strong vacuum force and care should be  

exercised to ensure that the suction and the position of the inlet  

end of the suction tubing is adjusted properly to prevent injury to  

the patient or inadvertent damage to surgical materials .

 

   The materials in the plume removed by this device are potentially 

hazardous . Handle according to 29 CFR 1910 .1030 and  

OSHA 3127 .1992 (Occupational Exposure to Bloodborne 

Pathogens) guidelines .

 

   To prevent a fire or explosion hazard, do not use the system in  

the presence of flammable or potentially flammable materials .

 

  Do not allow fluid to be pulled into the system .

 

   To prevent premature failure of the ULPA filter cylinder, do not  

operate this device without a disposable Pre-Filter in place .

 

   No modification of this equipment is allowed .

 

   No Customer serviceable parts .

 

   Only replacment of power cords or fuses can be done by the user . 

Only replace with these items with the exact replacement part(s) 

available from CooperSurgical .

Warning

To avoid risk of Electrical Shock, this equipment must only be connected to 

a supply mains with protected earth .

Содержание LEEP PRECISION LP-30-120

Страница 1: ...LEEP PRECISIONTM Smoke Evacuator 120 VAC Operating Manual REF LP 30 120 ...

Страница 2: ...r Use 2 INITIAL SETUP 2 ULPA FILTER INSTALLATION 2 OPERATING INSTRUCTIONS 2 Maintenance 3 Cleaning and Disinfection 3 Specifications 4 Reordering Information 5 Liability Statement 5 Warranty Information 6 To Obtain Warranty Service 7 EMC Compliance Information 8 Explanation of Symbols 12 i ...

Страница 3: ... removed by this device are potentially hazardous Handle according to 29 CFR 1910 1030 and OSHA 3127 1992 Occupational Exposure to Bloodborne Pathogens guidelines To prevent a fire or explosion hazard do not use the system in the presence of flammable or potentially flammable materials Do not allow fluid to be pulled into the system To prevent premature failure of the ULPA filter cylinder do not o...

Страница 4: ...able Pre Filter onto the ULPA Filter cylinder Photo B Be sure this device is firmly seated For procedures requiring close proximity plume removal i e vaginal speculum Assemble the inch Reducer Catalog number 6083 onto the port on the disposable Pre Filter top with a slight twisting motion Attach one end of an appropriate length of inch inner diameter ID Evacuation Tubing Catalog number 6084 to the...

Страница 5: ...equired during normal operation other than making sure that ample space is maintained around the unit to allow for free air flow and adequate cooling Cleaning and Disinfection After each patient use follow the combined cleaning disinfection instructions provided below for the LEEP PRECISION Smoke Evacuator CLEANING 1 Prepare an enzymatic neutral pH cleaner solution according to the Manufacturer s ...

Страница 6: ...t the wipes as necessary to ensure that all surfaces remain wet for the entire time 9 Allow the device to air dry Once dry visually inspect the device for any residual soil Repeat cleaning disinfection instructions if visible soil is present 10 Place the disinfected device in an appropriate dry storage area Specifications Use Environmental Temperature between 10 C and 40 C Relative Humidity betwee...

Страница 7: ...h and utilize natural resources prudently and rationally do not dispose of waste electrical or electronic equipment WEEE as unsorted municipal waste Contact local WEEE disposal sites Reordering Information Disposable Pre Filter Catalog REF number 52560 ULPA Filter Catalog REF number 6082 Reducer Catalog REF number 6083 Disposable Tubing inch ID Catalog REF number 6084 Liability Statement CooperSur...

Страница 8: ... by CooperSurgical nor normal wear and tear and does not apply to disposable or single or limited use items or components The sole and exclusive remedy under this Limited Warranty shall be repair or replacement as provided herein The foregoing Limited Warranty states the sole warranty made by CooperSurgical with respect to the Product and all parts thereof and is in lieu of any other warranty by C...

Страница 9: ...s repair covered under this Limited Warranty CooperSurgical will repair or replace the Product as provided above and will return the repaired or replacement Product to the customer freight prepaid with the risk of loss passing to the customer upon delivery to common carrier If CooperSurgical determines that the Product requires a repair that is not covered by this Limited Warranty CooperSurgical w...

Страница 10: ...ON Smoke Evacuator should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 CooperSurgical LEEP PRECISION Smoke Evacuators use only RF energy for their internal function Therefore RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class A...

Страница 11: ...ode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle ...

Страница 12: ...E1 P 80 MHz to 800 MHz 7 E1 P 800 MHz to 2 5 GHz 3 5 V1 P NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile...

Страница 13: ...imum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output rating of the transmitter in watts W according to the transmitter manufacturer For the CooperSurgical LEEP PRECISION Smoke Evacuator V1 3 Vrms E1 3 V m NOTE 1 At 80 MHz and 800 MHz the separation distan...

Страница 14: ...1 1 and CAN CSA C22 2 No 601 1 6D33 Date of Manufacture Explanation of Symbols Caution U S Federal law restricts this device to sale by or on the order of a physician RxOnly In order to preserve protect and improve the quality of the environment protect human health and utilize natural resources prudently and rationally do not dispose of waste electrical or electronic equipment WEEE as unsorted mu...

Страница 15: ...13 NOTES ...

Страница 16: ...37994 B Rev A 11 18 95 Corporate Drive Trumbull CT 06611 USA Phone 800 243 2974 Fax 800 262 0105 International Phone 1 203 601 9818 Fax 1 203 601 4747 www coopersurgical com Made in the USA ...