Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environmental Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2 .5 GHz
3 V
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the CooperSurgical
LEEP PRECiSION Smoke
Evacuator, including cables,
than the recommended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter .
Recommended separation
distance
d
=
d
=
d
=
where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d
is the recommended
separation distance in
meters (m) .
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
[ ]
3 .5
E
1
√
P
80 MHz to
800 MHz
[ ]
7
E
1
√
P
800 MHz to
2 .5 GHz
[ ]
3 .5
V
1
√
P
NOTE 1:
At 80 MHz and 800 MHz, the higher
frequency range applies .
NOTE 2:
These guidelines may not apply in all
situations . Electromagnetic propagation
is affected by absorption and reflection
from structures, objects, and people .
a
Field strengths from fixed transmitters, such
as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur
radio, AM and FM radio broadcasts, and TV
broadcasts cannot be predicted theoretically
with accuracy . To assess the electromagnetic
environment due to fixed RF transmitters, an
electromagnetic site survey should be considered .
If the measured field strength in the location in
which the CooperSurgical LEEP PRECISION
Smoke Evacuators used exceeds the applicable
RF compliance level above, the CooperSurgical
LEEP PRECISION Smoke Evacuator should be
observed to verify normal operation . If abnormal
performance is observed, additional measures may
be necessary, such as reorienting or relocating
the CooperSurgical LEEP PRECISION Smoke
Evacuator .
b
Over the frequency range 150 kHz to 80 MHz,
field strengths should be less than 3 V/m .