5
[Lst LOG]: permits to view the stored values on the display.
1.
Select the set-up procedure by pressing
MODE
and press
CAL.
2.
Select [Lst LOG] with
¿
or
À
and press
CAL
3.
The display shows the first data-logged value in memory (a blinking
[n] appears). Press
¿
or
À
to view all other values.
4.
While pressing
HOLD
the corresponding date is shown (not relevant
when a manual or continued data-logging has been performed).
5.
While pressing
RES
the corresponding time shown (or: log-
number, e.g. [2-017], when a continued data-logging has been
performed).
6.
While pressing °C the corresponding temperature is shown.
7.
Press
to send the recalled data to a printer.
8.
Press
MODE
to return to cancel.
Recalling data
[Cnt LOG]: Permits to continue with a next series of measurements,
without erasing the previous data-logging
1.
Select the set-up procedure by pressing
MODE
and press
CAL
.
2.
Select [Cnt LOG] with
¿
or
À
to continue with a next series of
measurements, without erasing the previous data-logging, provided
the memory has enough capacity left and press
CAL
.
Adding data
Set log: Procedure to pre-set the data-logging parameters.
1.
Select the set-up procedure by pressing
MODE
and press
CAL.
2.
Choose [SEt LOG] with
¿
or
À
. Select [no LOG] if no data-logging is
required and press
MODE
to return to the original mode.
3.
Press
CAL
untill you can choose with
¿
or
À
between [LOG PPM]
or [LOG %]. Select the desired mode and press
CAL
.
4.
The display shows e.g. [dt=100]. Select the desired time interval
(0...9999 s) between the data-logging, with
¿
or
À
and press
CAL
.
Pre-set to zero for manual data-logging, each time
is pressed
(the next points 5 to 7 are skipped and press
CAL
).
5.
The display shows e.g. [n=40]. Select the desired number of values
(1...100) to be data-logged with
¿
or
À
and press
CAL
.
6.
The display shows e.g. [d=31.12.91] while the year is blinking.
Program the desired starting date with
¿
or
À
and press
CAL
.
Program similarly the month and the day.
7.
The display shows e.g. [t=23.59] while the minutes are blinking.
Program the desired starting time with
¿
or
À
and press
CAL
.
Program similarly the hours.
8.
The display shows [StArt ?]. Press
CAL
to start the data-logging
according the previous settings (a blinking [L] appears) or press
MODE
to cancel. After the last value has been catalogued the display
will show [End LOG]. When manual data-logging has been selected,
press
to put a next measurement into memory. In the
meantime the display shows the logging-number e.g. [LOG 7]. Press
MODE
to cancel.
Logging data