4
[no LOG]
:
No data-logging.
[SEt LOG] :
Permits to pre-set the data-logging parameters.
[LSt LOG] :
Permits to view the stored values on the display.
[Prt LOG]
:
Permits to send the stored values to a printer or computer.
[Cnt LOG] :
Permits to continue with a next series of measurements,
without erasing the previous data-logging.
Up to 100 values can be stored in the non-volatile memory of the
instrument.
The measured values can be printed automatically at timed intervals or
manually on operator command.
Printer
1.
Select the set-up procedure by pressing
MODE.
2.
Press
CAL
until the display shows e.g. [Pt=120]. Select the desired
time interval (0...9999 s) between the transmitted data to a printer,
with
¿
or
À
. Pre-set to zero if no automatic printing is required.
3.
Press
CAL
to continue or press
MODE
to cancel.
The instrument can be identified by e.g. a computer when a specific
number is allocated to it.
Identification
number
Secret code
A private code can be programmed to avoid undesired access to the
instrument.
1.
Select the set-up procedure by pressing
MODE
.
2.
Press
CAL
until the display shows [Code on], [Code OFF], or [Code
CAL]. Decide, with
¿
or
À
, whether the access to all future
calibrations should be by entering a code (on) or not (off). Press
CAL
to continue or press
MODE
to cancel. Select [Code CAL] to enter
your personal code.
3.
The display shows [E=-----] while [=] is blinking. Enter your secret
sequence of 5 keys to proceed. Press
CAL
to continue or press
MODE
to cancel.
1.
Select the set-up procedure by pressing
MODE.
2.
Press
CAL
until the display shows e.g. [Id.no=5]. Select the desired
identification number (0...999) for your meter with
¿
or
À
. Pre-set to
zero if this number should not be mentioned while printing.
3.
Press
CAL
to continue or press
MODE
to cancel.
DATA-ACQUISITION