Alternating measurements
1.
Select the menu by pressing
MODE
.
2.
Select [ALTERNATE] and press
CAL
to activate or deactivate. The display
alternates between the selected modes with an interval of ca. 5 s.
3.
Press
MODE
to return to the measurements.
Good Laboratory Practice
1.
Press
CAL
while measuring.
2.
Select [GLP] and press
CAL
.
3.
Browse with
çèéê
to show a complete calibration report. Press
STORE
to
print the report.
Data Storage in the internal memory
1.
Select the menu by pressing
MODE
.
2.
Select [DATALOGGER] and press
CAL
.
Start the data-logging
1.
Select [STORE] and press
CAL
.
2.
Select [ACTIVATE] and press
CAL
.
3.
Select [YES] or [NO] and press
CAL
.
4.
Select [INTERVAL] and press
CAL
.
5.
Select the desired time interval between the data-logging (or MANUAL) and
press
CAL
.
6.
Select [ALARM] and press
CAL
.
7.
Select the desired time interval between the data-logging (or MANUAL), when
the alarm is activated, and press
CAL
.
8.
Select [NUMBER] and press
CAL
.
9.
Select the desired number of values to be data-logged and press
CAL
.
10.
Press
MODE
to return to the measurements and to start the data-logging
according to the previous settings. When manual data-logging has been
selected, press
STORE
to put a next measurement into memory. In the
meantime the display shows the logging-number e.g. [#00027].
View the stored values on the display
1.
Select [RECALL] and press
CAL
.
2.
Select [TABLE] or [PRINT] to display or print the stored data and press
CAL
to
continue. Follow the intructions on the screen.
3.
In case [TABLE] have been selected press
CAL
to show the options.
4.
Select [LOG#] and press
CAL
. Select the number of the first line to be shown
on the display and press
CAL
.
5.
Select [SHOW] and press
CAL
. Select [Log#] (line numbers) or [Date] to be
shown on the display and press
CAL
.
Erase the stored values
1.
Select [ERASE] and press
CAL
. Follow the intructions on the screen.
External storage in a computer using USB or RS232
1.
Select the menu by pressing
MODE
.
2.
Select [USB/RS232] and press
CAL
.
3.
Select [BAUDRATE] and press
CAL
.
4.
Select the desired baudrate and press
CAL
.
5.
Select [INTERVAL] and press
CAL
.
6.
Select the desired interval between the transmitted data and press
CAL
. Pre-
set to zero if no automatic transmitting is required.
7.
Select [ALARM] and press
CAL
.
8.
Select the desired interval between the transmitted data, when the alarm is
activated, and press
CAL
. Pre-set to zero if no automatic transmitting is
required.
9.
Select [IDENTIF. No.] and press
CAL
.
10.
Enter an identification number for the transmitted data and press
CAL
.
•
You can download free software and a suitable USB driver from
www.consort.be
.
4
Содержание T8710
Страница 1: ...Consort MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG T8710 T8720 April 2013...
Страница 2: ...T8710 16 channels for thermocouples type J K T or E T8720 12 channels for Pt100...
Страница 20: ...Specifications 17...
Страница 23: ...20...