Model VC940: OPERATING MANUAL
19
400mV
0.01mV
±(0.025%+8)
Around 2.5G
Ω
4V
0.0001V
40V
0.001V
400V
0.01V
1000V
0.1V
±(0.05%+5)
1000V
Around 10M
Ω
B. AC Voltage (AC+DC measurement is available)
Accuracy
Range
Resolution
Bandwidth
Accuracy
45Hz~1kHz
±(0.5%+5)
1kHz~10kHz
±(0.8%+8)
10kHz~120kHz
±(3%+8)
120kHz~250kHz
±(4%+8)
4V
0.0001V
250kHz~400kHz
±(10%+50)
45Hz~1kHz
±(0.5%+5)
1kHz~10kHz
±(1%+8)
10kHz~100kHz
±(3%+8)
40V
0.001V
100kHz~400kHz
Not Specified
45Hz~1kHz
±(0.5%+5)
1kHz~10kHz
±(4%+8)
400V
0.01V
10kHz~250kHz
Not Specified
45Hz~1kHz
±(1%+8)
1kHz~5kHz
±(5%+8)
750V
0.1V
5kHz~10kHz
±(10%+8)
Remarks:
z
Input Impedance: Approx 10M
Ω
.
z
Overload Protection: 1000V.
z
Display:
a)
True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0 except 750 where it is 1.5.
c)
A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d)
When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e)
When frequency is high than 100kHz, the accuracy gaurantee range 15%~100%
f) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading based on the above table.
C. DC Current
Range
Resolution
Accuracy
Overload Protection
400µA
0.01µA
±(0.25%+10)
4000µA
0.1µA
40mA
0.001mA
400mA
0.01mA
±(0.8%+10)
0.5A, 250V, fast type fuse,
ø5×20mm
10A
0.001A
±(1.5%+20)
10A, 250V, fast type fuse,
ø5×20mm
Remarks:
At 10A range:
z
When the measured current is
≤
5A, continuous measurement is allowed.
z
When the measured current is between 5A-10A, continuous measurement
≤
10 seconds and interval more than 15 minutes.
D. AC Current (AC+DC measurement is available)
Accuracy
Range Resolution
Bandwidth
Accuracy
Overload Protection
400µA
0.01µA
4000µA
0.1µA
40mA
0.001mA
400mA
0.01mA
45Hz~1kHz
1kHz~10kHz
±(0.5%+5)
±(1%+5)
0.5A, 250V, fast type fuse,
ø5×20mm
45Hz~1kHz
±(0.8%+5)
10A
0.001A
1kHz~10kHz
±(1.5%+5)
10A, 250V, fast type fuse,
ø5×20mm
Remarks:
z
Display:
a)
True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0.
c)
A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d)
When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e)
When frequency is high than 100kHz, the accuracy gaurantee range 15%~100%
f) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading based on the above table.
z
At 10A range:
a) When the measured current is
≤
5A, continuous measurement is allowed.
b)
When the measured current is between 5A-10A, continuous measurement
≤
10 seconds and interval more than 15 minutes.
E. Resistance
Range Resolution
Accuracy
Overload
Protection
400
Ω
0.1
Ω
±(0.8%+10)+test leads open circuit value
4k
Ω
0.001k
Ω
40k
Ω
0.01k
Ω
400k
Ω
0.1k
Ω
4M
Ω
0.001M
Ω
±(0.5%+10)
40M
Ω
0.01M
Ω
±(1.5%+20)
1000V
F. Continuity Test