Model VC940: OPERATING MANUAL
17
Chapter 6
Maintenance
This chapter provides basic maintenance information including battery and fuse replacement instruction.
Warning
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information.
A. General Service
z
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
z
To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings.
z
Turn the Meter to OFF when it is not in use.
z
Take out the battery when it is not using for a long time.
z
Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
B. Replacing the Fuses
Figure 6-1. Fuse Replacement
Warning
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure.
Follow Figure 6-1 and proceed as follows to replace the Meter’s fuse:
z
Turn the rotary switch to OFF and remove all connections from the terminals.
z
Remove the 5 screws from the case bottom.
z
Remove the fuse by gently prying one end loose, then take out the fuse from its bracket.
z
Install ONLY replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse, ø5×20mm
Fuse 2: 10A, 250V, fast type fuse, ø5×20mm
z
Rejoin the case bottom and case top, and install the 5 screws.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper operation.
C. Replacing the Battery
Figure 6-2. Battery Replacement
Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “
” appears.
Make sure the test leads are disconnected from the circuit being tested before opening the case bottom.
Follow Figure 6-2 and proceed as follows to replace the battery:
z
Turn the rotary switch to OFF and remove all connections from the terminals.
z
Remove the screw from the battery compartment, and separate the battery compartment from the case bottom.
z
Replace with a new 6F22 9V battery.
z
Rejoin the case bottom and battery compartment, and reinstall the screw.