4-5
4.3.7 Low Voltage Power Supply Module
Removal and Replacement
NOTE: This module is not user serviceable at
the component level. If faulty, the entire cir-
cuit board must be replaced. Replacements are
available from CONMED Customer Service.
Do not discard the module cover, mount-
ing plate and hardware; the replacement part
includes only the circuit board.
Low Voltage Power Supply Module Removal:
1) Remove Top Cover.
2) Loosen four screws located in slots and
unlatch the two cables.
3) Slide Low Voltage Power Supply Module
inward off the slots, then upward to remove.
NOTE: When servicing unit, board can be held in
heatsink as shown. This will provide access to the
lower boards while the unit is functioning.
4) Remove the cover by removing four screws.
Then remove the four standoffs to separate
the Low Voltage Power Supply from its
mounting plate.
Circuit Board
Low Voltage Power Supply Module Replacement:
1) Replace Low Voltage Power Supply on
mounting plate, fasten standoffs, replace cover
and tighten screws.
2) Place Low Voltage Power Supply Module
into unit on standoffs. Tighten the loose
screws and latch the two connectors.