Table 4.9 DIP Switch Settings
Config.
Switch
Position
Title /
Display
Element
Default Description for Off
Description for On
1
TEST /
Cut 100’s
Off
Run Mode. Required position for
surgery.
Activates Test Mode, which inhibits most
of the system level monitoring for trouble-
shooting purposes. When this switch
is ON, the Remote PC Switch on the
Display Panel must be pressed until the
Test Tone begins. If the Remote PC key is
not pressed, and Err 100 is displayed and
the power must be cycled.
2
CAL/ Cut
10’s
Off
Run Mode. Required position for
surgery.
Required for calibration of output power
and A.R.M.TM When this switch is ON,
the Remote PC Switch on the Display
Panel must be pressed until the Test Tone
begins. If the Remote PC key is not
pressed, and Err 100 is displayed and the
power must be cycled.
3
SCOAG /
Cut 1’s
On
Allows simultaneous activation
of two accessories for Pinpoint
Coagulation.
Allows activation of only one accessory for
Pinpoint Coagulation. Spray and Standard
Coagulation modes always allow simulta-
neous activation of two accessories.
4
DCUT /
Coag 100’s
Off
Allows activation of only one active
accessory in cut modes.
Allows activation of two active accessories
in cut modes.
5
LAST /
Coag 10’s
On
Defaults to Pure Cut, Pinpoint
Coag, and Macro Bipolar and sets
all power levels to zero (0W) each
time the system is initialized.
Defaults all modes and power levels to the
settings the last time the system was pow-
ered down.
6
PCEN /
Coag 1’s
Off
Allows the Remote Power Control
feature to be enabled by pressing
the PCEN switch on the Display
Panel.
Inhibits the Remote Power Control fea-
ture.
7
Not used
/ Bipolar
10’s
Off
Not used.
Not used.
8
DACview /
Bipolar 1’s
NA
DACview - See below.
DACview - See below.