INSTALLATION GUIDE
11/9/12
3DHD VISION SYSTEM
2
MARKETING COMMUNICATIONS
This installation guide is for use by ConMed Distributors to ensure proper set up of the
ConMed 3DHD Vision System. For more detailed information and for trouble shooting by
product please refer to the product User Manuals which can be found with each of the
individual 3DHD Vision System components.
Tools Required:
Utility Knife
Phillips Screw Driver
Country specific Power Cord for Cart
(outside the US, Canada, Mexico, and
Puerto Rico)
6MM Hex Bit Socket Wrench
Alignment Target Part #T05713-001
Tools Supplied:
4MM Allen Wrench
Inspect all shipping packages for any signs of mishandling or damage! In the event of
damage please contact Customer Service:
Email:
– 200 – 9824
CART:
To unpack cart cut strapping, remove top cover and slide side packaging up and
over cart.
Lift cart out of packaging base and place on floor.
Inspect for any damages, including cosmetic damages.
Phillips Screw Driver is a registered trademark of Phillips Screw Company
Allen in a registered trademark of Allen Manufacturing Company
Note:
If any discrepancies cannot be resolved please report immediately to Viking
Systems Technical Support!
[email protected] or call (800) - 200 - 8824
Note: Cart may or may not be shipped on its side, if so, cut strapping to release
cart from pallet and tip cart box to the upright position.
Note: Carts sent outside of the U.S. with the exclusion of Canada, Mexico and
Puerto Rico are shipped without an AC Power Cord. It is the responsibility of the
installer to provide the proper AC Cord.