INSTALLATION GUIDE
11/9/12
3DHD VISION SYSTEM
17
MARKETING COMMUNICATIONS
ALIGNMENT OF THE SCOPE/CAMERA:
The alignment should be checked periodically.
WARNING:
If the system comes
out of alignment it WILL cause eye strain and discomfort. In severe cases 3D
effect will be lost.
Do not wear
Viking System 3D glasses. Hold the tip of the Endoscope
approximately 4 to 5 inches (100
– 130mm) from the alignment target.
Targets
Included with Installation Guide.
Rotate the camera so that the long line of the cross is horizontal on the monitor.
Move the Endoscope closer to the target and ensure that the vertical lines of the
two crosses overlap.
Tip of the Endoscope should be approximately 2 inches (50mm) from the target.
The horizontal lines shall also overlap so that there is no visible gap between them,
as shown in
Figures 41 and- 42
.
Figure 41 -
GOOD
Figure 42 -
BAD
IF MISALIGNMENT IS DETECTED:
Remove the Endoscope from the camera head.
Direct the camera head at the target, keeping its distance from the target
approximately 20 inches (500mm). Observe two images of the target.
Rotate the camera head so that the long line of the cross is made horizontal on the
monitor.
Make sure both lines from the two images are horizontal and overlap. NOTE: The
vertical line will not overlap by design. If the horizontal lines are misaligned as
shown
in figure 44
the camera head came out of alignment, most likely having
been dropped. If the camera head is aligned then the scope is the issue, see
figure 43
.
Figure 43 -
GOOD
Figure 44 -
BAD