INSTALLATION GUIDE 

 

11/9/12 

3DHD VISION SYSTEM 

 

16 

 

MARKETING COMMUNICATIONS 

 

 

 

  Once white balance completes the on screen display (OSD) looks (for a brief 

period) as shown is 

Figure 39

 

 
 
 
 
 
 
 
 

 

 
 
                                             

Figure 39   

 

  Press and hold the white button on the 3D camera head.  The white balance 

should initiate and complete as displayed above. 

  Press and release the left Blue Button on the camera head and the control unit will 

beep once.  Now press and hold the same left Blue Button and the control unit will 
beep twice.  The left Blue Button as shown in 

Figure 40

 

 

                             

Figure 40 

 

  Press  and  release  right  Blue  Button  and  the  brightness  on  screen  display  (OSD) 

will  appear  on  the  monitor.    Now  when  pressing  left  and  right  buttons  on  the 
camera head while the brightness is displayed the curser on screen will move left 
and right and the video signal brightness on monitor will change. 

 

 
 

 

 

WHITE BALANCE 

 

OK 

 
 

MODE LAPAROSCOPY 1  

Содержание 3DHD

Страница 1: ...by copyright Copying disclosure to others or sue of this publication is prohibited without express written consent of ConMed Corporation ConMed Corporation reserves the right to make changes show her...

Страница 2: ...stomer Service Email customerservice vikingsystems com or phone 800 200 9824 CART To unpack cart cut strapping remove top cover and slide side packaging up and over cart Lift cart out of packaging bas...

Страница 3: ...t b Plug the Hospital Grade Power Cord into the outlet under the green pilot light of Power Strip Secure the Hospital Grade Power cord using the clip provided with the Video Cart as shown in Figure 3...

Страница 4: ...Video Cart place the right side of the Mounting Bracket Power Supply under the right side of the 2nd shelf with the cord to the Sony Monitor facing you b Fold the Mounting Bracket Power Supply up so i...

Страница 5: ...h cart channel as shown in Figures 7 10 Figure 7 Figure 8 Figure 9 Figure 10 Attach one of the 1 m IEC 320 power cords to power supply as shown in Figure 11 Figure 11 Open 3DHD controller packaging an...

Страница 6: ...t column covers Unpack Sony 3D monitor Inspect for any damages including cosmetic damages Vesa Mounting pattern on back of the Sony monitor once unpacked should look like picture shown in Figure 15 Fi...

Страница 7: ...shown in Figure 19 Figure 19 Connect left black HD SDI video cable to Sony 3G Card input 1 Connect right white HD SDI video cable to Sony 3G Card input2 MONITOR AT THIS POINT WILL LOOK LIKE FOLLOWING...

Страница 8: ...ng tilt swivel VESA mount plate and remove the existing VESA mounting plate See Figure 21 2 Using the four 4 existing screws install the 200 200 VESA mounting plate to the existing mounting holes prov...

Страница 9: ...deo cable to the SDI 2 input as shown in Figure 25 Figure 25 THE MONITOR AT THIS POINT WILL LOOK LIKE THE FOLLOWING PICTURE IN FIGURE 26 Figure 26 3DHD CONTROLLER Unpack the 8170 6 3DHD controller Ins...

Страница 10: ...light source Inspect for any damage including cosmetic damages Place the light source on top of the 3DHD controller position feet in detents on controller cover As shown in Figure 28 Figure 28 Attach...

Страница 11: ...h right and left eyes The on screen display OSD should only be present in the left eye The on screen display OSD should not be visible with right eye only Unpack 8170 9 Dual Channel Camera Head or 817...

Страница 12: ...asses and verify the double image displayed horizontal edges are at the same horizontal plane as shown in Figure 33 Figure 33 STEREO ENDOSCOPE Unpack the 8191 05 06 Dual Channel Stereo Endoscope or 81...

Страница 13: ...light guide Inspect the outer sheathing for nicks cuts or voids Aim one end of light guide up at the overhead lighting and view other end to verify light passing through you can cover and uncover end...

Страница 14: ...Stand By button to allow the light source to emit light to the light guide as shown as shown in Figure 36 Figure 36 The tip of the Endoscope should be emitting light WARNING Do not aim light directly...

Страница 15: ...nd aim 3DHD Camera Head with Scope and Light Guide attached at a non reflective white object as shown in Figure 37 Figure 37 Press White Balance Button on front of camera controller and verify on moni...

Страница 16: ...ue Button on the camera head and the control unit will beep once Now press and hold the same left Blue Button and the control unit will beep twice The left Blue Button as shown in Figure 40 Figure 40...

Страница 17: ...e should be approximately 2 inches 50mm from the target The horizontal lines shall also overlap so that there is no visible gap between them as shown in Figures 41 and 42 Figure 41 GOOD Figure 42 BAD...

Страница 18: ...ING IF PURCHASED WITH SYSTEM Remove 8170 7 2D camera head from its packaging and inspect the camera housing and cable for cuts tears cracks or other defects Connect camera head to control unit Attach...

Страница 19: ...rol unit will beep twice To ensure you are achieving full 2D resolution aim the camera head at an image and verify wearing the Viking Systems 3D glasses that an image is present in the left and right...