PC-POD-CP-006-v01
PC-POD-CP-006-v01
Printed copies are not controlled.
Page
9
of
35
7.1.1.5
Stick the band electrode around the participant at the first attachment
location, as shown below, making sure that the folded back ends of the
tape are joined on the anterior surface of the participant.
NOTE: Use enough tension so that the band sticks, but not too much to the point
that it wrinkles or twists.
7.1.1.6
Measure at least 3cm from the lower edge of the conductive part of
electrode band #1 (metallic/colored band) and mark the location.
7.1.1.7
Using the electrode tape, measure the quantity required by placing the
upper edge of the conductive part of the electrode tape on the mark,
leaving an extra 3-4 inches at each end of the electrode tape and cut
accordingly.
7.1.1.8
Place the tape on a flat surface with the metallic/colored part facing
upward.
7.1.1.9
Strip off the electrode tape insulation backing.
7.1.1.10
Apply conductive electrode gel to the metallic/colored surface on
the sticky side of the tape leaving 3-4 inches at each end not coated.