PC-POD-CP-006-v01
PC-POD-CP-006-v01
Printed copies are not controlled.
Page
10
of
35
7.1.1.11
Fold-back the ends of the tape so that the sticky part is not
exposed.
7.1.1.12
Stick the band electrode tape around the participant’s neck so
that the mark is aligned with the upper edge of the conductive part of
the electrode tape. Make sure that the folded back ends of the tape are
joined on the anterior surface of the participant.
7.1.1.13
Place electrode band #3 directly under the breast line (under
xyphoid process) following the same procedure as described above.
NOTE: For female participants, make sure that the electrode band is not in contact
with the participant’s bra.
7.1.1.14
Measure at least 3cm from lower edge of the conductive part of
electrode band #3 (metallic band) and mark the location.
7.1.1.15
Place electrode band #4 so that the mark is aligned with the
upper edge of the conductive part of the electrode tape following the
same procedure as described above.
7.1.1.16
Measure the distance in cm between the lower edge of the
conductive portion of band #2 and the upper edge of the conductive
portion of band #3 in both the anterior and posterior of the participant
and take note for future use.
7.1.1.17
Connect the appropriately numbered electrode clip of the DB-9
connector to both folded ends of the folded back ends of the electrode
band.
7.1.1.18
Once placed, tape the alligator clip down to secure, if needed.