PC-POD-CP-006-v01
PC-POD-CP-006-v01
Printed copies are not controlled.
Page
18
of
35
8.12
Click “OK”.
8.13
On ICG machine, press “OP” key to switch back to operate mode, which allows
data to be collected.
9.
Data Acquisition
9.1
Opening Acquisition Window
9.1.1
From the main menu select “Acquisition” then “Data Acquisition and Setup”.
This window will allow for operational parameters to be specified.
9.1.2
Under “Acquisition Mode” select the desired mode of data acquisition.
Options are the following:
9.2
OPTION 1 – Collect data intermittently using inter-sample pauses
9.2.1
For acquisition of impedance data intermittently at preset time intervals select
“Continuous Time Intervals (Auto)” under Acquisition Mode. Enter the
desired “Inter-Sample Interval Time” in either minutes (blue) or seconds (red).