When using a TrueFeed for the first time, do an initial
loadcell calibration
as follows:
The unit must be mounted horizontally (water-leveled)
Avoid vibrations during the loadcell calibration. This will influence the calibration.
Do not touch the unit during loadcell calibration.
When using the optional slide frame, the whole unit must be slid completely in against the throat adaptor
and fixed.
Be sure that the TrueFeed loadcell is connected to the controller
Go to the LOADCELL menu
In this menu it is possible to calibrate the loadcell of the TrueFeed
(500gr. Calibration weight required)
Select “YES” to start the loadcell calibration
[***] = Progress
After a few minutes the following screen appears:
Place calibration weight (500 gr.) on the hopper
and press the “START” button
After approximately 1 minute the loadcell calibration is ready, press the “MENU” button to continue.
To check if the loadcell calibration was OK, go to the WEIGHT CHECK menu.
><:
Standstill sign. When the vibrations are too big, this sign disappears!
Weight: Actual weight on the weighing scale (gram)
Object: Object
weight
(gram)
Zero:
Zero YES / NO. Reset the object weight.
‘Zero’ the object weight
Place 500 gr. calibration weight on the hopper
If the object weight is not corresponding with the real weight, perform another loadcell calibration.
LOADCELL CALIBRATION
[***]
Status
Calibration busy.
Please wait…………..
LOADCELL CALIBRATION
[***]
Status
Place weight, press
START when ready.
WEIGHT CHECK
Weight: gr. ><
Object: gr.
Zero: NO/YES
22 Dual Feed TrueFeed Manual