Maintenance and cleaning
40
–
Absorbent cloth
–
Bonding agent
–
Protection caps
Procedure
Release coupling nipples from the coupling sockets.
Secure the coolant inlet and coolant outlet connections with protection caps against
soiling.
The coolant hoses are now disconnected from the unit.
9.5. Connecting the coolant hoses
NOTE
Risk of damage by using improper or faulty coolant hoses!
This may lead to damage to persons, damage to property or corrosion damage.
When choosing coolant hoses pay attention to sufficient burst strength and compatibility with
coolant.
Only use coolant hoses without any signs of damage.
In case water is being used as coolant, ensure that non-transparent hoses are used to prevent
algae growth in the water. Otherwise appropriate additives have to be used.
NOTE
When connecting the coolant hoses pay attention to flow direction. Follow the documentation re-
leased by the manufacturer of the unit to be cooled.
Transparent coolant hoses stimulate algae growth that may increase the error-proneness of the
components built into the unit. Thus, only use non-transparent coolant hoses.
The coolant hoses are connected to the unit by means of couplings. Coolant inlet and coolant out-
let are labeled with respective symbols.
Figure 9: Labeling of coolant inlet and coolant outlet
Connecting the coolant hoses
Requirements
Unit prepared for maintenance, see "Preparing the unit for maintenance".
Procedure
Remove the protection caps from the coolant inlet and coolant outlet connections of
the unit.