8
IE-SNC-T3110-09
Info mode
Several settings of installed transmitter are possible to verify without a use
of the computer. It is necessary to connect at least power of current loop I1.
Unscrew the transmitter lid and shortly press button between display and
interconnection terminals by means of a tool (e.g. screwdriver).
The LCD display shows the type of value assigned
to 1. output (here
„%RH“ = relative humidity and „1“
= loop I1). Upper line displays value of the current
corresponding to measured value on the lower line
(here 4 mA corresponds to 0 %RH).
Press button again to get value for upper point of
the 1. output (here
„%RH“ = relative humidity and 1“
= loop 1). Upper line displays value of the current
corresponding to measured value on the lower line
(here 20 mA corresponds to 100 %RH).
After next press of the button LCD display shows
the type of value assigned to
2. output (here „
ºC
“ =
temperature and
„2“= loop I2). Upper line displays
value of the current corresponding to measured
value on the lower line (here 4 mA corresponds to -
30
ºC).
Press button again to get value for upper point of
the 2.
output (here „ºC“ = temperature and „2“ =
loop 2). Upper line displays value of the current
corresponding to measured value on the lower line
(here 20 mA corresponds to 80
ºC).
Press button again to end info mode and display actual measured values.
Warning
During info mode no measurement and no output current generation
proceed. The transmitter stays at info mode 15 s and then automatically
goes back to measuring cycle.
The verifying of settings of the T3110Ex transmitter is permitted only in
non-hazardous area.