Absolute humidity:
Accuracy:
±1.5 g/m
3
at ambient temperature T < 25
°C,
for more details see graph
Measuring range:
0 to 400 g/m
3
Specific humidity
This value depends on atmospheric pressure. Pressure for quantities
calculation is stored in device memory. Default value is 1013hPa and can
be changed by software.
Accuracy:
±2 g/kg at ambient temperature T < 35 °C
Measuring range:
0 to 550 g/kg
Mixing ratio
This value depends on atmospheric pressure. Pressure for quantities
calculation is stored in device memory. Default value is 1013hPa and can
be changed by software.
Accuracy:
±2 g/kg at ambient temperature T < 35 °C
Measuring range:
0 to 995 g/kg
Specific enthalpy
This value depends on atmospheric pressure. Pressure for quantities
calculation is stored in device memory. Default value is 1013hPa and can
be changed by software.
Accuracy:
±4 kJ/kg at ambient temperature T < 35 °C
Measuring range:
0 to 995 kJ/kg
The values computed from ambient temperature and relative humidity
including their accuracy you can exactly determine by the program
Conversions