SECT IO N 6
-
TEST LEVEL CALIBRATION
Rev050321
Set the frequency of the signal generator to 150 kHz (without
modulation).
Adjust the signal generator output to
either
:
the respective
SGout
value determined during calibration for the
present test frequency.
-or-
the level at which the power indicated by the power meter is equal to
the respective
FWDpwr
value determined during calibration for the
present test frequency
.
Record the amplitude measured on the measuring instrument as
U
mr
.
Increase the signal generator output by 5.1 dB, and record the
amplitude measured on the measuring instrument as
U
mr(inc)
. The
difference between
U
mr
and
U
mr(inc)
must be between 3.1 dB and 7.1
dB.
Increase the frequency of the signal generator to the next test
frequency.
Repeat steps 2 through 5 until the next frequency in the sequence
would exceed the highest frequency of the test.