TEST LEVEL CALIBRATION
Rev050321
On a conductive ground plane, connect the EM Clamp and Calibration
Accessories/Adapters as shown in Figure 17. The bottom surface of the EM Clamp,
and both 150Ω to 50Ω Adapters should be flush against the top surface of the
ground plane, and should be either:
a) fastened using bolts or screws directly to the ground plane (through their
respective mounting holes); or,
b) bonded to the ground plane using copper tape with conducted adhesive
(3M #1181 HD recommended).
The ADA-515-CLEM
150Ω to 50Ω Adapter is a non
-inductive 100 ohm resistor as
defined in the IEC 61000-4-6 standard.
The TEP-
050 50Ω Termination is a 2.5W, 50 ohm termination.
Distributed by: Reliant EMC LLC, +1 408 916
‐