INSTRUCTION MANUAL
SECT IO N 6
-
TEST LEVEL CALIBRATION
Rev050321
The frequency range of the test is, in most cases, 150 kHz to 80 MHz; and, in
some cases, 150 kHz to 230 MHz. The calibrations and tests are performed
at discrete, logarithmically spaced frequencies (as opposed to a
frequency sweep). The logarithmic spacing of the test frequencies means
that the actual step size increases after each step. The maximum step size
at any given frequency is equal to 1% of the present frequency, as
calculated using Equation 7.
A truncated example of the test frequencies, showing the respective step
sizes is shown in Table 5.