SEC TI ON 5
-
CALIB RATION & TE ST PROCED URE S - MI L-STD-461G (CS114)
Rev081219
1) Set up the equipment as described in the applicable standard or approved
test plan. An example setup diagram is given in Figure 14.
2) Turn on all equipment and allow sufficient time for stabilization. Set the
frequency of the signal generator and measurement equipment to 10 kHz,
unmodulated.
3) Slowly increase the amplitude of the signal generator until the value indicated
on Spectrum Analyzer/EMI Receiver #1 reaches the appropriate amplitude
(see section 5.2).
4) Record the present frequency and the measured forward power value
indicated on Spectrum Analyzer/EMI Receiver #2.
5) Set the frequency of the signal generator and measurement equipment to the
next test frequency (see section 5.1).
6) Repeat Steps 3-5 until all test frequencies have been calibrated.
NOTE:
Typical CS114 Forward Power Levels for the CLCI-400 are shown in Figure 8, in
Section 3.7 of this document.
Содержание CLCI-400
Страница 2: ...Page 2 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 3: ...Page 3 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 4: ...Page 4 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 6: ...Page 6 of 30 INSTRUCTION MANUAL SECTION 2 PRODUCTS AVAILABLE FROM COM POWER Rev081219...
Страница 12: ...Page 12 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 13: ...Page 13 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 14: ...Page 14 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 15: ...Page 15 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...