Page 23 of 30
INSTRUCTION MANUAL
SEC TI ON 5
-
CALIB RATION & TE ST PROCED URE S - MI L-STD-461G (CS114)
Rev081219
1) Set up the equipment as described in the applicable standard or approved
test plan. An example setup diagram is given in Figure 16. Refer to MIL-STD-
461G or the applicable test plan to determine the EUT cables to be tested.
2) Turn on all EUT equipment and test equipment and allow sufficient time for
stabilization. Set the frequency of the signal generator and measurement
equipment to 10 kHz with 1 kHz pulse modulation, 50% duty cycle. Verify that
the modulation is present on the drive signal and that modulation frequency,
waveform and depth (40 dB minimum from peak to baseline) are correct.
3) Without exceeding the maximum current level for the applicable limit (refer to
Table 6 below), adjust the output level of the signal generator until the forward
power value indicated on Spectrum Analyzer/EMI Receiver #2 is equal to the
respective forward power value recorded during Step 4 of the Test Level
Calibration Procedure described in Section 5.3.
Содержание CLCI-400
Страница 2: ...Page 2 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 3: ...Page 3 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 4: ...Page 4 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Страница 6: ...Page 6 of 30 INSTRUCTION MANUAL SECTION 2 PRODUCTS AVAILABLE FROM COM POWER Rev081219...
Страница 12: ...Page 12 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 13: ...Page 13 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 14: ...Page 14 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Страница 15: ...Page 15 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...