INSTRUCTION MANUAL
SEC TI ON 5
-
CALIB RATION & TE ST PROCED URE S - MI L-STD-461G (CS114)
Rev081219
There are five severity levels at which the CS114 test can be performed (Curves 1
through 5). Refer to Table VI of the appropriate MIL-STD-461 document to
determine the appropriate curve for your application. Curves 1 through 5 are
defined in Figure CS114-1, and in Figure 13 below.
As shown in Figure 13, each curve begins at 10 kHz, ascending linearly with the
logarithm of frequency at a rate of 20 dB/decade until it reaches 1 MHz. Each
curve remains at a constant amplitude between 1 MHz and 20 MHz; and then
ascends linearly with the logarithm of frequency at a rate of 8 dB/decade until it
ends at 200 MHz.
The value at any point on either slope can be calculated using the values and
formulas shown in Table 5.