5. Instructions for Use
5.28 Opening the Door in Case of Power Failure
If there is a power failure, remove the kickplate of the HYDRIM. Remove the screw
with a screwdriver and lift the kickplate up and toward you. Locate the ring and pull
it to open the door. Exercise caution! There may be fluid remaining in the unit and
the instruments may be hot. Instruments that have not completed the cycle should
not be used and should be reprocessed. When replacing the kickplate, ensure that
the grounding wire is attached.
Kickplate
Screw
26
Grounding wire
Ring