6. Care and Maintenance
6.7 Annual maintenance
If the message “Preventative Maintenance Required. Call for Service” appears, the annual
maintenance is required. To continue using the HYDR
IM
, press OK. Call for service as soon
as possible. The annual maintenance should be done after 365 days have elapsed since the
initial installation or since the last service call was performed.
6.8 Unit testing and validation
The following outlines the recommended regular maintenance to be performed by the user
and technicians for the HYDR
IM
C61wd G4 washer-disinfector.
To ensure the unit's ongoing cleaning efficacy, SciCan recommends observing the following
periodic testing protocols.
IMPORTANT NOTES:
This document outlines the minimum requirements to ensure that the equipment performs
to specification. In some regions, local/national guidance documents may require additional
testing. Please contact your dealer or local medical authorities for details.
•
Installation, commissioning, annual servicing
MUST
be undertaken by SciCan approved
technicians. Failure to maintain this equipment may invalidate the results of the following
testing regime.
The recommended periodic testing protocols are listed in the table below:
When
Who
What Tests
Upon installation
By a
SciCan-approved
technician
•
Commissioning/Installation testing as
outlined in the installation document included
with the unit.
•
First validation if required by local regulations.
Daily tests and checks
By the user
•
door lock check
•
wash arm rotation check
•
door seal check
•
load carrier check
•
check and clean chamber filters
•
visual examination (inspection under
magnification) of each load for residual soil.
Annually
By a
SciCan-approved
technician
•
Annual maintenance schedule
as outlined in the service manual
Annually
By a
SciCan-approved
technician
•
Re-validation if required by local regulations.