background image

Page 15

3.  Setting up your HYDR

IM

3.16  Adjusting the button beep volume

If you would like to adjust the beep volume, follow these steps:

1. 

2. 

Scroll to 

Beep Volume

 and select.

3. 

Use 

 to scroll through the volume settings. Select the one you want by  

pressing it. Press 

 to save and move back to the Setup menu. 

3.17  Adjusting the salt regeneration

Salt regeneration should be set according to the local water hardness. See section 3.4 
Testing Water Quality for instructions on determining correct settings. To set salt  
regeneration, follow these steps:

1. 

2. 

Scroll to 

Set Regeneration

 and select.

3. 

Use 

 to change the value. The default setting is 1. Press 

 to save and return 

to the Setup menu. 

3.18  Adjusting the screen contrast

The touchscreen is calibrated for the lighting condition of most sterilization centers. Should 
you need to adjust the contrast for your office, follow these steps:

1. 

2. 

Scroll to 

LCD Contrast

 and select.

3. 

Use 

 to scroll through your contrast options. When you have found the contrast 

you require, press it. Press 

 to save and return to the Setup menu. 

3.19  Changing the touchscreen display themes

The touchscreen themes (i.e. icons and background colours) can be changed to one of the 
preset options. To change themes follow these steps:

1. 

2. 

Scroll to 

Theme

 and select.

3. 

In the

 

Change Theme

 

screen, use 

 to scroll through your available options.  

As you scroll, each theme will display on the touchscreen. Press 

 to select your 

theme and return to the Setup menu.

Содержание SciCan HYDRIM C61 wd G4

Страница 1: ...l rights reserved I SO1588 3 H Y D R I M C 61wd Fully C o m p l i a n t Operator s Manual Benutzerhandbuch USB Manual de uso USB Manuale operatore USB Manuel de l utilisateur USB USB USB Podr cznik ob...

Страница 2: ...H info eu scican com Wymagany jest czytnik formatu PDF Aby otrzyma dokument w formie papierowej nale y skontaktowa si z firm SciCan GmbH pod adresem info eu scican com PDF l sare kr vs F r en pappersk...

Страница 3: ...escription charts Baskets and accessories Hints and tips for best cleaning results Recommended load configurations Running a cycle Interrupting aborting a cycle Additional user and setup menu function...

Страница 4: ...the property of their respective owners Contents of this manual are subject to change without notice to reflect changes and improvements to the HYDRIM product The HYDRIM should only be installed and s...

Страница 5: ...than that specified the protection provided by the equipment may be impaired Cleaning solutions may irritate Avoid contact with eyes skin and mouth Never sit stand or lean on the open door The unit ma...

Страница 6: ...1 HEPA filter n 1 Dryer exhaust n 1 Door latch n 1 n 1 USB port n 1 LCD touchscreen n 1 Dryer outlet n 1 Water softener n 1 Wash arms top not shown n 1 Coarse filter n 1 Cleaning solution n 1 and dra...

Страница 7: ...Page 6 2 Important Information Settings Unlock door Water softener detergent level and door status Network connectivity green when active USB status green when active 2 3 Touchscreen overview Programs...

Страница 8: ...een Instructions Instructions Delay Remote Access Prepare for Shipping Set Printer Language Country Date Time Unit No Network Setup Reset Drying Counter Reset LCS Counter Screen Saver Temperature C F...

Страница 9: ...g installation If the HYDRIM C61wd G4 is installed in a sterilization center the manufacturer of the sterilization center should allow enough space at the top back and both sides of the unit to facili...

Страница 10: ...ersonal protective equipment you use for clinical activities to handle HIP products Always ensure that personal protective equipment if contaminated with HIP Ultra cleaning solution is cleaned or disp...

Страница 11: ...ny of these spots or stains a water test may be required to determine the cause The installation of a water treatment system may be required to reduce the amount of dissolved solids in the water and t...

Страница 12: ...dd 0 5 kg of water softening salt to the water softener container using the supplied funnel to prevent any salt from spilling into the chamber and close by screwing the lid tightly back into place An...

Страница 13: ...ut its policies and programs governing the disposal of electronic devices 3 6 Setting the language The messages displayed by your HYDRIM can be presented in a number of different languages To change t...

Страница 14: ...me and select Date Setup 3 From the DATE screen use the keypad to set the date Press EN to save and to return to the Setup menu 4 To change the format in which the date appears return to the Setup men...

Страница 15: ...When you have found the amount of time you require press it Press to save and return to the Setup menu 3 14 Adjusting the temperature display 1 2 Scroll to Temperature C F and select 3 Use to choose...

Страница 16: ...is 1 Press to save and return to the Setup menu 3 18 Adjusting the screen contrast The touchscreen is calibrated for the lighting condition of most sterilization centers Should you need to adjust the...

Страница 17: ...a User PIN From the User PIN screen you can assign up to four PINs To assign a PIN follow these steps 1 2 Scroll to User and select 3 To assign a user PIN select User PIN and use the numeric keypad t...

Страница 18: ...s a 10 100Base T Ethernet port located at the back of the unit To connect your HYDRIM to a network using a router follow these steps 1 Connect your network cable to the Ethernet port at the back of th...

Страница 19: ...n your network speed and making an initial connection can take longer 3 24 Connecting to a wireless network HYDRIM models offer Wireless capability for connecting units to the G4 network adding flexib...

Страница 20: ...the HYDRIM include Stainless steel scalers spatulas and hemostats Stainless steel retractors needle holders skin hooks tissue forceps scissors curettes and other stainless steel instruments free of de...

Страница 21: ...e equivalent exposure time to a specified temperature Cycle Prewash Wash Rinse Dry P0 Machine Cleaning Cycle No initial draining 30 C cold 2 10 minutes default 2 minutes N A 30 C cold 2 minutes N A P1...

Страница 22: ...vities should be placed so that water can drain freely Large basket 01 113545 Handpiece adapter 01 113625 Hygiene basket 01 113546 Two basket rack 01 113257 Rack for Vertical and Hinged Instruments Re...

Страница 23: ...nstruments and ensure that water can flow unrestricted in and out of the tubes Do not place suction tubes horizontally in a basket or in a cassette For instruments containing lumens such as dental han...

Страница 24: ...the unit Handle all instruments with care to prevent personal injuries from punctures for example and follow local health and safety regulations to prevent accidents and injuries Using baskets and acc...

Страница 25: ...ts solid instruments and open hollow instruments For example Tissue forceps and scissors For example Scalers curettes dental elevators single part mirrors suction tubes etc For example Bone files plug...

Страница 26: ...decrease depending on the projected cycle completion 8 The cycle is complete When the cycle is complete the program name and Cycle Complete is displayed and the unit beeps three times Press OK to open...

Страница 27: ...TE If the drying phase of the cycle is interrupted the critical parts of the cycle are complete and the message displaying the program name and Cycle Complete Drying interrupted will appear 4 8 Additi...

Страница 28: ...ons Users can access a database of pictorial instructions on how to perform routine maintenance such as installing the cleaning solution 4 9 Opening the door in case of a power failure If there is a p...

Страница 29: ...on 2 The unit will record up to the last five successful cycles and the last five incomplete cycles If you select a cycle from the list it will display cycle information in a format similar to how it...

Страница 30: ...from the web portal 3 When the activity light on your USB storage device stops blinking or the USB icon on the LCD turns from a flashing green to a solid grey remove the USB storage device and transf...

Страница 31: ...eaning solution NOTE To avoid spillage replace the solution container only when the message appears and be sure to wear gloves When a red X appears on the water softener detergent icon Press on the ic...

Страница 32: ...mark will appear in place of the red X when it is ready for use NOTE The system can also be primed by simply starting a cycle and selecting Detergent Replaced when prompted Place a new bag in the chem...

Страница 33: ...ilter and bacteriological filter if fitted Air filter When the message Replace air filter appears the HEPA filter is in need of changing 1 Turn the power off 2 Pull open the ventilation panel below th...

Страница 34: ...he arrow mark on the filter matches the direction of the arrow on the bracket Push the right hand filter fitting into tube B 4 Gently press the replacement filter into the filter bracket The arrow mar...

Страница 35: ...run a chamber cleaning cycle Press OK to continue the selected wash cycle but remember to run a chamber cleaning cycle at the next available opportunity 6 6 Preparing the unit for shipping If prepari...

Страница 36: ...ional guidance documents may require additional testing Please contact your dealer or local medical authorities for details Installation commissioning annual servicing MUST be undertaken by SciCan app...

Страница 37: ...dealer for technical service Load carrier check Visually check load carriers trolleys for wear and tear damage and that the runners move freely over their full range Chamber filter check Remove visual...

Страница 38: ...ntinued protection against the risk of fire replace only with 15A 250V type F fuses The wash arms do not rotate Remove the wash arms clean and reinstall them Check for obstructions in chamber The door...

Страница 39: ...ounter See section 3 12 Air Filter Missing Install Filter Check HEPA filter to ensure it is installed correctly with arrow in correct position Lost USB storage device with all cycle data If you need t...

Страница 40: ...from the unit Check that the drain hose is not kinked Check that the filters in the bottom of the chamber are not obstructed Check that the drain outlet is within the specified height noted in the in...

Страница 41: ...that water hoses Hot Cold are not reversed Restart cycle CF15 Water Reservoir Overflow Water reservoir overflow switch triggered Restart program Call for service CF16 Ambient Temperature Error Unit i...

Страница 42: ...eed too low Check that load is not blocking wash arm Check all holes in wash arms for debris If message persists call for service CF24 Bottom wash arm low RPM Bottom wash arm speed too low Check that...

Страница 43: ...invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan s warranties and other duties with respect to the quali...

Страница 44: ...mperature 60 C 140 F Drain 3 4 Drying system Heater 1 kW Electrical connection 200 230VAC 10 single phase 50 Hz 12A 60 Hz 12A Protection class Class I Equipment pollution degree Pollution degree 2 Equ...

Страница 45: ...arked as confidential or proprietary at the time of disclosure c Control will mean the possession directly or indirectly of the power to direct or cause the direction of the management and operating p...

Страница 46: ...ovide Customer with at least six 6 months advance notice Customer will allow SciCan to use remote access software to aid in the resolution of problems or questions Update fees if any will be billed on...

Страница 47: ...release of the SciCan Software Product or b the combination operation or use of the SciCan Software Product with hardware programs or data not furnished or otherwise approved by SciCan if such infrin...

Страница 48: ...of Updates or by persons other than SciCan employees or consultants 7 4 DISCLAIMER THE WARRANTIES ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPL...

Страница 49: ...t disclosed or distributed by its employees or consultants in violation of the provisions of this Article 9 The parties agree that the terms and conditions of this Agreement are considered confidentia...

Страница 50: ...ing upon written notice to SciCan Customer may assign or otherwise transfer this Agreement to an Affiliate of Customer provided such Affiliate agrees with SciCan to be bound by the terms and condition...

Страница 51: ...ht or any other right in the future 10 10 Amendment This Agreement may be amended only by a written document executed by a duly authorized representative of each of the parties 10 11 Exclusive Agreeme...

Страница 52: ...mum 0 111W RF Output power Extensible Authentication Protocol While the HYDRIM unit is connected to a WiFi network the security of the connection depends on the configuration of the Wireless infrastru...

Страница 53: ...r network it may be worth checking that it s configured for the highest level of security If your network is configured for an older generation of security WEP or WPA Wi Fi Alliance recommends you mov...

Страница 54: ...s Setup The HYDRIM unit allows for Wireless or Wired connections but only one type at a time 2 3 1 Selecting between the Wired and Wireless networks 2 3 2 Connecting to a Wireless network 11 WiFi Regu...

Страница 55: ...requirements The device has been tested and found to comply with the limits for a Class B digital device pursu ant to FCC Part 15 Subpart B Due to differences in channel allocation if you cannot conn...

Страница 56: ...device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can...

Страница 57: ...erference received including interference that may cause undesired operation Responsible Party SciCan LTD 1440 Don Mills Road Toronto Ontario Canada M3B 3P9 Telephone 1 800 667 7733 5 2 Industry Canad...

Отзывы: