Owner’s Guide
7
WARNING!
To reduce the risk of water-borne illness:
•
Flush water supply to this device regularly.
Electromagnetic Interference
• Radiation interferes with operation of
implanted medical devices such as pacemakers,
causing chest pain and shortness of breath.
•
Use of this device adjacent to or stacked with
other equipment should be avoided because
it can result in improper operation.
•
Use of accessories and cables other than those
provided by the manufacturer of this device
could result in increased electromagnetic
emissions or decreased electromagnetic
immunity of this device and result in improper
operation.
Use only accessories provided by COLTENE
intended for use with this equipment.
Do not use in oxygen rich environments.
VII. Precautions for Ultrasonic Prophylaxis
Procedures
• Ultrasonic inserts “wear out” with use. Inserts that have 2 mm
of wear lose about 50% of their scaling efficiency.
• If excessive wear is evident, or the insert has been bent, discard
the insert immediately. It may break during use.
• Use retraction to protect the tongue, cheek and lips when the
scaler is in use to avoid contact between the two.
VIII. Order Information
Units:
Catalog No.
Voltage
Frequency
Power
US100R115M
115 V
50/60 Hz
100W max
US100R115T
115 V
50/60 Hz
100W max
US100R230CE 230 V
50/60 Hz
100W max
US100R230UK 230 V
50/60 Hz
100W max
US100R240
240 V
50/60 Hz
100W max
Inserts: Frequency 25 kHz
Catalog No.
Description
Qty
US1025K #10 Universal
1
USG1025K #10 SuperSoft
1
US1025KOM #10 OptiMist™
1
US1025KSP
#10 Universal Slim
1
USG1025KSP
#10 Slim SuperSoft
1
US1025KOS
#10 Slim OptiMist™
1
USG1025KOS
#10 Slim SuperSoft (Optimist™)
1
US100025K
#1000 Universal Triple Bend
1
Inserts: Frequency 30 kHz
Catalog No.
Description
Qty
US1030K #10 Universal
1
USG1030K #10 SuperSoft
1
US1030KOM #10 OptiMist™
1
US1030KSP
#10 Universal Slim
1
USG1030KSP
#10 Slim SuperSoft
1
US1030KOS
#10 Slim OptiMist™
1
USG1030KOS
#10 Slim SuperSoft (Optimist™)
1
US100030K
#1000 Universal Triple Bend
1
US10OR (OR-alpha)
Replacement O-Ring Gaskets
1
US100RXF
Water Filter Replacement Kit
1
IX. Troubleshooting
Generally, check all lines and connections to and from the unit,
a loose plug or connection will often create problems. Check
the settings on the unit’s knobs. Although service and repair of
the BioSonic
®
US100R ultrasonic scaler should be performed by
authorized personnel, the following are some basic trouble shooting
procedures that will help avoid unnecessary service calls.
A. Unit will not operate
1. Check the ON/OFF switch making sure it is in the ON position,
and that the detachable power cord is fully seated in the
receptacle on back of unit (ON/OFF indicator light should
be on).
2. Check that the unit’s electrical plug is fully seated in an
appropriate AC receptacle, and that the receptacle is active.
3. Disconnect the power cord and check the fuse(s).