Service Menu VIBA X.30
2021-05-05 Seite 17/21
5.6.2
TEST ENDPOS
TEST ENDPOS
900
OPEN
CLAMP
EDIT
EDIT ENDPOS
900
+
-
Set mixing unit end stop.
OPEN
= Test upper end stop
CLAMP
= Test lower end stop
EDIT
= Modify ENDPOS parameter
The mixing unit is opened and closed via a DC motor the power
consumption of which is monitored.
When the end stop is reached the power drawn by the motor
exceeds the set value of the current threshold (ENDPOS) and the
clamping motor is shut off.
The factory set parameter for the end stop has been selected so
that the mixing unit arrives directly at the end position without a
prior switching off.
Before modifying the parameter check whether the threaded
spindles are dirty or damaged. If necessary, clean and apply grease
to the spindles as described in the operating instructions.
A larger value for the DC-MOT ENDPOS means a higher current
intake and thereby a larger force of the clamping motor.
The value must not be set too high, otherwise the threaded spindle
could jam.
IMPORTANT:
The ENDPOS parameter does not influence the clamping force;
the clamping force is set in the menu item 5.5 SET CLAMP FORCE