Service Menu VIBA X.30
2021-05-05 Seite 11/21
5.3.3
EDIT SPEED PGM
EDIT PGM-1
Total
: 0:40
Time (0) : 0:10
Speed (0) :
40%
+
-
Setting of time for each programme step.
+
= Increase speed for programme step 0
-
= Reduce speed for programme step 0
Key E
= Quit submenu
The speed can be set at 40 to 100% of the maximum rotation speed.
The area from 1 to 39% must not be used as this, depending on the
mixing can, can lead to resonance phenomena.
If 0% is selected the mixing unit will not move. A pause can be
programmed in this way.
IMPORTANT:
A mixing speed which is too high can cause damage to the mixing
can. Check the suitability of the mixing can through a mix test after
each speed modification.
5.4
SET BUTTONS
SET BUTTONS
T1 T2 T3
I
II
III
The assignment of mixing times T1 -T3 (preset under 5.2 SET TIME
SET TIME) to keys I-III can be set in the SET BUTTONS menu.
T1 -T3
= The mixing time can be selected directly
AUTO
= The mixing time is selected based on the con-
tainer height
BLANK
= The key is not assigned to a function
The desired assignment can be changed by pressing the keys I-III.
IMPORTANT:
With AUTO, the mixing time is automatically selected depending
on the container height.