INSTRUMENT SETTINGS
These allow setting of temperature logs, auto restart and date and time.
TEMPERATURE LOG
When set to
ON
the temperature log will record the block temperatures for every run and save them as
a Report. If temperature logs are not required, set to
OFF
.
AUTO-RESTART
This determines whether the unit will restart after a power failure. Setting Auto-restart to
Always
will
allow the unit to re-start irrespective of the length of the power failure. Selecting
Never
will mean the
system will not auto-restart after a power outage. In addition, the user can select a time in which, if
the power comes back on in that window, then the system will resume. The options are shown below.
Note:
the time periods in the auto-restart allow the end user to select windows of power loss which
they deem acceptable for the power to be disrupted during their protocols. Example: if the system is
set to 10 minutes and the power is out for 8 minutes the system will resume, if the power is out for 12
minutes the system will not auto-restart.