background image

4

OPERATING INSTRUCTIONS

The motor runs at a constant speed of 1725 rpm. The pump speed
setting is changed by using the lever on the mechanical speed control.
To increase the pump speed, move the lever away from the mechani-
cal gear box, to decrease pump speed, move the lever toward the gear
box. The pump speed can be varied from approximately 10 to 400 rpm.
The knob on the lever locks the setting and prevents accidental changes
in the pump speed (rotate the knob CW (clockwise) to lock the setting
and CCW (counterclockwise) to unlock).

To minimize the starting load, it is recommended that the system be at
approximately zero speed (lever fully CW) prior to starting the motor.
Once the motor has started, move the lever to increase speed to the
desired setting and lock in place.

Operating the system continuously at speeds approaching zero will
cause excessive wear on the mechanical speed control. Operating at
the maximum speed will result in minimum tubing life. If maximum sys-
tem life at a continuous speed is desired, the size of the pump head
should be chosen so that the desired flow is achieved at a pump speed
of approximately 60 to 200 rpm. For intermittent operation, any pump
head can be used at any speed setting without harm.

FIGURE 2

Untitled-1

2/17/04, 7:27 AM

8

Содержание Masterflex 7583-50

Страница 1: ...DUTY PUMP DRIVES Cole Parmer Instrument Co 625 East Bunker Court Vernon Hills Illinois U S A 60061 1844 800 MASTERFLEX 627 8373 847 247 2929 Fax www masterflex com e mail techinfo coleparmer com A 12...

Страница 2: ...Untitled 1 2 17 04 7 27 AM 2...

Страница 3: ...ION 1 SETUP AND DRIVE OPERATION 2 Wiring 2 GENERAL INSTRUCTIONS 3 OPERATING INSTRUCTIONS 4 PUMP MOUNTING 5 Add On Head Capacity 6 MAINTENANCE 6 Cleaning 6 Service Parts 6 Lubrication 6 SPECIFICATIONS...

Страница 4: ...which may be present WARNING To prevent ignition of hazardous atmospheres this system must not be installed in an area where vapor or gases having an ignition temperature less than 280 C are present C...

Страница 5: ...MASTERFLEX L S and MASTERFLEX I P Pump Heads with standard short shafts Ad ditional MASTERFLEX Pump Heads with standard short shafts may be used in tandem The pump speed can be varied from 10 to 430 r...

Страница 6: ...plicable local codes The motor must be wired such that the output shaft of the MOTOR rotates in a clockwise direction when facing the motor shaft See the wiring diagram below 2 WIRING DIAGRAM For Mode...

Страница 7: ...ting plate opening Then select the appropriate adapter for the pump you wish to drive and slip it into the rubber sleeve and seat it firmly see Figure 2 The pump mounting plate which is mounted to the...

Страница 8: ...minimize the starting load it is recommended that the system be at approximately zero speed lever fully CW prior to starting the motor Once the motor has started move the lever to increase speed to t...

Страница 9: ...ERFLEX L S and MASTERFLEX I P pumps All of the MASTERFLEX pumps mount in a vertical orientation See Figure 3 Orient the tang of the pump you wish to drive with the drive adapter and install the pump t...

Страница 10: ...L S Drive Adapter B 3545 Rubber Coupling Sleeve B 3434 Coupling End 3 8 in Bore Speed Reducer B 3546 Coupling End 5 8 in Bore Motor B 3433 Coupling End 3 4 in Bore Motor European B 3676 Lubrication T...

Страница 11: ...nce Painted steel Operating Voltage Frequency 115V AC nominal 57 63 Hz 90 130V AC 4 7 A AC Power Rating 1 4 hp 190W Output Speed Range 10 to 430 rpm Direction of Rotation Clockwise Speed Regulation 10...

Страница 12: ...duct misuse or abuse may be invoiced to you The warranty period for this product is noted on the Warranty Card PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for au thor...

Отзывы: