Page 24
USING
THE
SERVICE
PARAMETERS
F4
–INITIAL ZERO RANGE
To enter this parameter, press the
[
➔
0/T
]
key when
“F4 int”
is shown.
The display will show the current initial zero range.
Press the
[Pcs/ ]
key to change the value and press
[
➔
0/T
]
to accept the value.
Press
[Print/M+/esc]
to return to weighing.
F5 -RE-ZERO RANGE
To enter this parameter, press the
[
➔
0/T
]
key when
“F5 rEZ”
is shown.
The display will show the current re-zero range.
Press the
[Pcs/
]
key to change the value.
Press
[
➔
0/T
]
to accept the value.
Press
[Print/M+/esc]
to return to weighing.
F6 -SUCCESSIVE TARE
To enter this parameter, press the
[
➔
0/T
]
key when “
F6 SCS”
is shown.
The display will show if the successive tare is on or off.
Press the
[Pcs/
]
key to change the value.
Press
[
➔
0/T
]
to accept the displayed value.
Press
[Print/M+/Esc]
to return to weighing.
F7
–ADC COUNTS
To enter this parameter, press the
[
➔
0/T
]
key when
“F7 Cnt”
is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the
[
➔
0/T
]
key to return to the PARAMETER menu.
Press the
[Print/M+/Esc]
key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.).
Typical value at full capacity is 500,000 (approx.).
F8
–ZERO MODE
To enter this parameter, press the
[
➔
0/T
]
key when “
F8 ZEm
” is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used. The other 2
zero modes are for unique locations in the world and effect the +/- range of the zero.
Press the
[Pcs/
]
key to change the value.
Press
[
➔
0/T
]
to accept the displayed value.
Press
[Print/M+/Esc]
to return to weighing.