Page 12
•
Press the
[
➔
0/T
]
key. “
P1 PCt
” will be displayed.
•
Press
[
➔
0/T
]
again to enter percent weighing. The scale will now accept a value
the user enters as the reference weight.
•
Enter the value using the
[Unit/
]
key and
[Pcs/
]
key to select and set the value of
the 100% mass.
•
When set, press the
[
➔
0/T
]
key to store the value.
The display will show “
0.00 %
”.
•
Press
[Unit/
]
to return to weighing.
ANIMAL
(DYNAMIC)
WEIGHING
The scale can be set to animal (dynamic) weighing for weighing any items that are unstable
or may move. See Section 7.4.
The scale will use a special filter to minimize the effects of any movement while the moving
animal or unstable samples are on the scale.
•
Press
[Func/C]
. The first option is “
FunC 1
”, press the
[Func/C]
key again to display
“
Func 2
”.
•
Press the
[
➔
0/T
]
key. “
F1 PCt
” will be displayed. Press the
[Pcs/
]
key to advance
to the second function, “
P2 Ani
”, Animal weighing.
•
Press
[
➔
0/T
]
to enter animal weighing. The scale will change to animal weighing
and allow weighing of an unstable animal or sample on the platform.
•
To use the Animal Weighing function it is necessary to set the amount of filtering
required for the item to be weighed. The display will show
“Flt x”
where
x
is a value
from 1 to 5. The higher the value the greater the amount of filter will be. More active
animals will require a higher level of filtering to give a stable result. To increment the
value shown, press the
[Pcs/
]
key then press the
[
➔
0/T
]
key to accept it.
•
The display will flash “
Ani
“
2
times then show the current weight, 0.00. and the
Animal/Dynamic weighing symbol will appear.
The scale is now ready to
weigh.